FR1MF1-B(NC150 BT)

EN

NOTE:

Patients and thermometer should stay in similar room 

condition for at least 30 minutes.

Don't take a measurement while or immediately after nursing a 

baby.

Don't use the thermometer in high humidity environments.

Patients should not drink, eat or exercise before/while taking the 

measurement.

Don't move the measurement device from the measuring area 

before hearing the termination beep.

10 short beeps and a red LCD backlight alert the patient that he/

she may have a temperature equal to or higher than 37.5 °C.

Always take the temperature in the same location, since 

temperature readings may vary according to locations.

Doctors recommend rectal measurement for newborn infants 

within the first 6 months, as all other measuring methods might 

lead to ambiguous results. If using a non contact thermometer on 

those infants, we always recommend verifying the readings with a 

rectal measurement.

In the following situations it is recommended that three temper-

atures are taken with the highest one taken as the reading:

1. Children under three years of age with a compromised 

immune system and for whom the presence or absence of 

fever is critical.

2. When the user is learning how to use the thermometer for the 

first time until he/she has familiarized himself/herself with the 

device and obtains consistent readings.

3. If the measurement is surprisingly low.

Readings from different measuring sites should not be 

compared as the normal body temperature varies by 

measuring site and time of day

, being highest in the evening 

and lowest about one hour before waking up.

Normal body temperature ranges:

- Axillar: 34.7 - 37.3 °C / 94.5 - 99.1 °F

- Oral: 35.5 - 37.5 °C / 95.9 - 99.5 °F

- Rectal: 36.6 - 38.0 °C / 97.9 - 100.4 °F

- Microlife FR1MF1-B(NC150 BT): 35.4 - 37.4 °C / 95.7 - 99.3 °

F

Changing between Celsius and Fahrenheit

This thermometer can display temperature readings in either Fahr-

enheit or Celsius. To switch the display between °C and °F, simply 

turn OFF the unit, 

press and hold

 the START button 

4

 for 5 

seconds; after 5 seconds, the current measurement scale («

°C

» or 

«

°F

» icon) will flash on the display 

AP

. Change the measurement 

scale between °C and °F by pressing the START button 

4

. When 

the measurement scale has been chosen, wait for 5 seconds and 

the unit will automatically enter the «ready for measuring» mode.

9. How to recall 30 readings in Memory Mode

This thermometer can recall the last 30 readings with a record of 

both time and date.

Recall mode 

AQ

Press the M-button

8

 to enter recall mode 

when the power is off. The memory icon «

M

» will flash.

Reading 1 - the last reading 

AR

Press and release the M-

button

8

 to recall the last reading. Number «1» and a flashing

«

M

» are displayed.

Reading 30 - readings in succession: 

Press and release the 

M-button

8

 consecutively to recall the last 30 readings in 

succession.

Pressing and releasing the M-button

8

 after the last 30 readings 

have been recalled will resume the above sequence from reading 1.

10.Bluetooth

®

 Function

This device can be used in conjunction with a smartphone running 

the «Microlife Connected » App. The measurement results 
will be automatically transferred via Bluetooth

®

.

Downloading the «Microlife Connected » App

Download «Microlife Connected » App for free from Google 

Play™ (Android) or App Store (iOS) and install it on your smartphone.

How the Bluetooth

®

 Function works

The Bluetooth

®

 function on your device will automatically turn on 

and will be ready to connect with the «Microlife Connected » 

App after the device is switched on. Your device will automatically 

upload the data once it is connected to the smartphone.

Bluetooth

®

 Icon Indicator

The Bluetooth

®

 icon indicator on your device, located in the middle 

left portion of the display, is designed to provide information about 

the connection between your device and smartphone.

NOTE:

The Bluetooth

®

 function is enabled: The Bluetooth

®

 icon indi-

cator 

AK

 appears steady on the display.

Содержание FR1MF1-B

Страница 1: ...crolife USA Inc 1617 Gulf to Bay Blvd Floor 2 Clearwater FL 33755 UNITED STATES Toll Free Help Line 1 800 568 4147 www microlifeusa com Made in China Microlife FR1MF1 B NC150 BT 2a 2b 3 FR1MF1 B NC150...

Страница 2: ...n the home Table of Contents 1 The Advantages of this Thermometer Measures in a matter of seconds Multiple uses wide range of measurement Accurate and reliable Gentle and easy to use Multiple readings...

Страница 3: ...s device connects to the Microlife Connected Health App by using Bluetooth 4 0 and enables easy monitoring of temperature 2 Important Safety Instructions Follow instructions for use This document prov...

Страница 4: ...ically switches to ready for measuring AK Cancel time setup Press the ON OFF button 5 during time setup The LCD will show Date Time icons with After that press the ON OFF button 5 to start the measur...

Страница 5: ...readings in either Fahr enheit or Celsius To switch the display between C and F simply turn OFF the unit press and hold the START button 4 for 5 seconds after 5 seconds the current measurement scale...

Страница 6: ...device in water or other cleaning liquids Take care not to scratch the surface of the sensor lens and the display 13 Battery Replacement This device is supplied with 2 new long life 1 5V size AAA batt...

Страница 7: ...5 IEC 60601 1 IEC 60601 1 2 EMC IEC 60601 1 11 ISO 80601 2 56 Expected service life 5 years or 12000 measurements This device complies with part 15 of the FCC Rules Operation is subject to the followi...

Страница 8: ...his device fulfills the stipulations of the standard IEC 60601 1 2 EMC Technical alterations reserved According to the Medical Product User Act a biennial technical inspection is recommended for profe...

Страница 9: ...PR 11 Group 1 The model FR1MF1 B NC150 BT uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipm...

Страница 10: ...ercial or hospital environment Surge IEC 61000 4 5 1 kV line s and neutral N A Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and v...

Страница 11: ...ectromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and...

Страница 12: ...ted maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 23 0 1 0 38 0 73 1 1 2 2 3 10 3...

Отзывы: