Guidance and manufacturer’s declaration – electromagnetic immunity
T
he model FR1MF1-B(NC150 BT) is intended for use in the electromagnetic environment specified
below. The customer or the user of the model FR1MF1-B(NC150 BT) should assure that it is used in
such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
N/A
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the model
FR1MF1-B(NC150 BT), including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
Distance in metres (m).b
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,a
should be less than the compliance level in each
frequency range. b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
te
lephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the model FR1MF1-B(NC150 BT) is used exceeds the applicable RF
compliance level above, the model FR1MF1-B(NC150 BT) should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating the model FR1MF1-B(NC150 BT).
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.