Microlife FR1MF1-B Скачать руководство пользователя страница 12

Recommended separation distances between

portable and mobile RF communications equipment and the model 

FR1MF1-B(NC150 BT)

 

The model FR1MF1-B(NC150 BT) is intended for use in an electromagnetic environment in which 

radiated RF disturbances are controlled. The customer or the user of the model FR1MF1-B(NC150 BT) 

can help prevent electromagnetic interference by maintaining a minimum distance between portable 

and mobile RF communications equipment (transmitters) and the model FR1MF1-B(NC150 BT) as 

recommended below, according to the maximum output power of the communications equipment. 

Rated maximum 
output power 
of transmitter 

Separation distance according to frequency of transmitter 

150 kHz to 80 MHz 

80 MHz to 800 MHz 

800 MHz to 2.5 GHz 

0.01 

0.12 

0.23 

0.1 

/

 

0.38 

0.73 

/

 

1.2 

2.3 

10 

/

 

3.8 

7.3 

100 

/

 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended separation 
distance d in metres (m) can be estimated using the equation applicable to the frequency of the 
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to 
the transmitter manufacturer. 
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects and people. 

3.

Содержание FR1MF1-B

Страница 1: ...crolife USA Inc 1617 Gulf to Bay Blvd Floor 2 Clearwater FL 33755 UNITED STATES Toll Free Help Line 1 800 568 4147 www microlifeusa com Made in China Microlife FR1MF1 B NC150 BT 2a 2b 3 FR1MF1 B NC150...

Страница 2: ...n the home Table of Contents 1 The Advantages of this Thermometer Measures in a matter of seconds Multiple uses wide range of measurement Accurate and reliable Gentle and easy to use Multiple readings...

Страница 3: ...s device connects to the Microlife Connected Health App by using Bluetooth 4 0 and enables easy monitoring of temperature 2 Important Safety Instructions Follow instructions for use This document prov...

Страница 4: ...ically switches to ready for measuring AK Cancel time setup Press the ON OFF button 5 during time setup The LCD will show Date Time icons with After that press the ON OFF button 5 to start the measur...

Страница 5: ...readings in either Fahr enheit or Celsius To switch the display between C and F simply turn OFF the unit press and hold the START button 4 for 5 seconds after 5 seconds the current measurement scale...

Страница 6: ...device in water or other cleaning liquids Take care not to scratch the surface of the sensor lens and the display 13 Battery Replacement This device is supplied with 2 new long life 1 5V size AAA batt...

Страница 7: ...5 IEC 60601 1 IEC 60601 1 2 EMC IEC 60601 1 11 ISO 80601 2 56 Expected service life 5 years or 12000 measurements This device complies with part 15 of the FCC Rules Operation is subject to the followi...

Страница 8: ...his device fulfills the stipulations of the standard IEC 60601 1 2 EMC Technical alterations reserved According to the Medical Product User Act a biennial technical inspection is recommended for profe...

Страница 9: ...PR 11 Group 1 The model FR1MF1 B NC150 BT uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipm...

Страница 10: ...ercial or hospital environment Surge IEC 61000 4 5 1 kV line s and neutral N A Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and v...

Страница 11: ...ectromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and...

Страница 12: ...ted maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 23 0 1 0 38 0 73 1 1 2 2 3 10 3...

Отзывы: