15
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery
current is consumed. We recommend having
batteries in the monitor even with mains operation,
just as a back-up. They ensure that in the event of
a power failure date and time are retained and do
not have to be entered manually again.
8. Error messages
When an error occurs during a measurement, the
measurement is interrupted and an error message
displayed, e.g. «Err 3».
If you do not receive an error message, but
if the results seem unusual to you, please
check whether you observed all points of the
checklist on page 8 and in the enclosed flyer
“8 golden rules of taking the blood pressure ”.
For more detailed help in troubleshooting the
monitor, please refer to the download section
of the www.aponorm.de product website.
Error
Designa-
tion
Potential cause and
remedy
«ERR 1»
16
-E
Signal too
weak
The pulse signals on
the cuff are too weak.
Re-position the cuff and
repeat the measure-
ment.*
«ERR 2»
16
-C
Error signal
During the measurement,
interference signals
were detected at the
cuff, caused for instance
by movement or muscle
tension. Repeat the
measurement, keeping
your arm still.
«ERR 3»
16
-D
Irregular
cuff pres-
sure
It is not possible to build
up sufficient pressure in
the cuff. There may be
a leak in the cuff. Check
that the cuff connection
is correct and that the
cuff is not too loose
around your arm. Replace
the batteries if necessary.
Repeat the measurement.