AQUADYN® FZ50 Series Hollow Fiber Module | OPERATION MANUAL
MICRODYN-NADIR | REVISION: B | DATE: 01/08/2020
14
Operating data
– this is where the
operating flux is determined
(1) Clean water flux before chemical
cleaning
Pure water counter pressure cleaning
(2) Measurement of flux
Cleaning with mixed chemical solution
(3) Measurement of flux
Cleaning with sodium hypochlorite
water solution
(4) Measurement of flux
(2) ≥ 90% of operating flux
(3) ≥ 90% of operating flux
(4) ≥ 90% of operating flux
End
Yes
Yes
Yes
No
No
No
Figure 1.
By measuring the flux between each cleaning step, it is easy to determine if further or more intense
cleaning is necessary in order to restore membrane performance.