Nabi User Manual
There is Standard Curve page in Bradford, Lowry, BCA, Biuret, and Others under
the Protein Measurement. It is necessary for the Standard Curve to be set before
performing the measurements. Standard Curve Panel can be opened by touching
Standard Curve button. There are 5 slots for measurements of samples with different
concentrations in order to generate Standard Curve. However, the Standard Curve will be
shown after measurement of 3 samples. Users can also save, open, and reset the
Standard Curve data by touching the Save, Open, and Reset buttons.
Measurement
Loads Measurement page. It is a default page when the users enter
any pages under Protein.
Standard Curve
Loads Standard Curve page for a Standard Curve setting. Users can
set, load a Standard Curve for the measurement, or save current
Standard Curve data.
Conc. (ng/µl)
The concentrations(ng/µl) of the sample 1 to 5.
Read
Measures to achieves Abs. Values of Sample1 to 5.
Abs.
Shows the Abs. Values of Sample1 to 5.
Reset
Reset all the values in the Standard Curve Panel.
Wavelength
Shows the wavelength of the current measurement.
Open
Opens the Open panel of saved Standard Curve data.
Save
Saves the current Standard Curve data and its settings.
Reset
Deletes current Standard Curve data.
Bradford Standard Curve Page
17
3. User Interface