28
Connect 18K ohm resistor at input and try to match the measurement value by pushing CAL +
or CAL – button. Then push C-20k button to advance to next procedure. The following screen
shows:
Connect 180K ohm resistor
Connect 180K ohm resistor at input and try to match the measurement value by pushing CAL +
or CAL – button. Then push C-200k button to advance to next procedure. The following screen
shows:
Connect 1.8 Meg ohm resistor
Connect 1.8Meg ohm resistor at input and try to match the measurement value by pushing CAL
+ or CAL – button. Then push C-2meg button to advance to next procedure. The following
screen shows:
Apply 1.8VDC
To stop calibration, push Calibration button.
This is the end of calibration. To restart a new calibration procedure, then follow the previous
procedures.