Page 53
Error Codes
scanCONTROL 30xx
12.
Error Codes
(— LED
Error
lights for a long time, • LED
Error
lights briefly)
Flashing
sequence
Cause
Remedy
Notes
Group: Loading / saving configuration
..
2x short
Mode not found.
Select different one.
Only previously stored
modes can be called up.
..-
2x short, 1x long
White error flash
Contact manufacturer, return
sensor.
Should not occur in normal
operation.
…
3x short
Flash full
None, contact manufacturer.
Should not occur in normal
operation.
….
4x short
Loading suppressed due to
active data transmission.
Stop active data transmission. Prevents PC software
crashes.
Group: Data processing and transmission
--
2x long
Data overflow in the sensor
Select smaller measuring field,
reduce profile frequency, se-
lect less complex measuring
program.
Data can be impaired; ex-
posure time can be longer
than expected.
--.
2x long, 1x short
Data overflow during receipt
of the data from the sensor.
Select smaller measuring field,
reduce profile frequency, se-
lect less complex measuring
program.
Data can be impaired.
--..
2x long, 2x short
Data overflow for serial port
RS422
Reduce profile frequency, se-
lect less complex measuring
program.
Data can be impaired.