Page 36
Operation of the Sensor with a PC
scanCONTROL 30xx
The number of rows evaluated on the sensor matrix is the set union of the covered rows of region of interest 1
and region of interest 2.
If the sensor is used in high-speed mode, only every second row is used for evaluation and thus the number
of covered rows by the ROIs (region of interest) is divided by two.
- Number of points per profile for processing. Following figures are valid:
2048 points per profile: Max. Frequency = 2703 Hz
1024 points per profile: Max. Frequency = 4000 Hz
512 points per prfile: Max. Frequency
= 7143 Hz
256 points per profile: Max. Frequency = 10000 Hz
The measuring field can be restricted by omitting complete matrix areas in order to suppress interfering im-
age ranges.
Measuring field and measuring range must be clearly differentiated in practical use. The measuring field is
related to the sensor matrix and the measuring range is related to the measuring object (the object space).
Both do not have to match on account of the optical mapping and the definitions.
The scanCONTROL 30xx sensors have the following characteristics
- The receiver has a smaller opening angle (view angle) than the laser line.
- Centered measuring field (symmetrical to the center axis).
- The high resolution sensor image matrix has 2048 x 1088 pixels. The measuring field geometry is fixed.
- Reference for the distance (Z-axis) is the lowest body edge of the sensor, see Chap.
- Use of the GigE-Vision standard. Standard GigE Vision implementation from different manufacturers can
be used.