Page 8
scanCONTROL 2910/2960BL
9.
Quickstart: Initial Operation, Software
Mount the sensor according to the installation instructions.
Install the software.
You will find the software Online on
the product website of the sensor
or in the download area:
https://www.micro-epsilon.
de/2D_3D/laser-scanner/Software/
downloads/
.
Follow the dialog instructions through the
installation process:
A. Read the installation help
B. Install the software
C. Additional information in the Online docu-
mentation
Connect the sensor to the PC using the optionally available SCR3000A-x Ethernet-
cable.
Connect the shield of the power supply cable to the PE grounding conductor of
the main power supply. Close connection plugs you do not require with the includ-
ed protective caps. Turn on the power supply (24 VDC).
Please wait until the scanCONTROL measuring system is recognized by the PC.
This may take a few seconds.
Now you can operate the scanCONTROL measuring system with the scanCONTROL
software packages.
i
Operating the sensors via Ethernet may require the IP settings of the PC/sensor
or the firewall settings to be adjusted, see Chap. 5.2.5 of the operating instruc-
tions.
The connectors for Ethernet and the multifunction port include a screw connector.
Loosen the screws connection before you remove the connector from the socket.
i
Only connect the sensor to the peripheral equipment, if it is disconnected from
the power supply, i.e. only when the supply voltage is switched off. The sensor
needs a warm-up time of typically 20 minutes for high precision measurements.