Page 24
Operation
MSC7401 Miniature sensor controller for inductive sensors
5.3.3
Sensor Parameter
LED D2 = red flashing
With this function, you can set the parameters
- Sensor model
- Excitation frequency and
- Excitation amplitude,
if the automatic recognition is not successful, or for special areas of use other settings may be necessary.
These depend on the used sensor model. After manual setting of the sensor parameters, the adjustment of
the system is recommended, see Chap.
5.3.4
.
Sensor model
Measuring range Sensor type Frequency Amplitude
DTA-1x
±1 mm
LVDT
5 kHz
550 mV
DTA-3x
±3 mm
5 kHz
DTA-5x
±5 mm
5 kHz
DTA-10x
±10 mm
2 kHz
DTA-15x
±15 mm
1 kHz
DTA-25x
±25 mm
1 kHz
LDR-10
10 mm
LDR
21 kHz
LDR-25
25 mm
13 kHz
LDR-50
50 mm
9 kHz
LDR-14
with 8 mm drawbar
14 mm
tbd
with 10 mm drawbar
tbd
LVP-25
with 8 mm drawbar
25 mm
tbd
with 10 mm drawbar
16 kHz
Fig. 16 Sensor models and sensor parameters