A 3.3.9.5 TRIGGERSW, Software Trigger Pulse
TRIGGERSW
Creats a trigger pulse. Error message is displayed if “SOFTWARE” is not selected in trigger selection.
A 3.3.10
Interfaces
A 3.3.10.1 BAUDRATE, RS422
BAUDRATE 9600|19200|56000|115200|128000|230400|256000|460800|691200|921600|
1000000
Set the baud rate for the RS422 interface.
A 3.3.10.2 UNIT, Web Interface
UNIT MM|INCH
Change the measurement display on the websites. The command has no effect on the ASCII interface.
- MM representation in mm
- INCH representation in customs
A 3.3.10.3 MFIFUNC, Function Selection Multifunctional Input
MFIFUNC NONE | MASTER | TEACH | TRIGGER
Choose function of the multifunctional input.
- NONE: multifunctional input has no function
- MASTER: multifunctional input is master impulse input
- TEACH: multifunctional input is teach input for analog output
- TRIGGER: multifunctional input is trigger input
A 3.3.10.4 ERROROUT1, Activate Digital Output
ERROROUT1 NONE|DIST|TEACH|LI1
Choose error signal of the digital output.
- NONE: error ourput deactivated
- DIST: no peak found or beyond measuring range (out of range)
- TEACH: Distance is out of scaled analog range
- LI1: Distance is greater than the limit value (ERRORLIMIT)