background image

External laser switching

Series LLT27xx(502), LLT27xx(503), LLT27xx(504) and LLT27xx(505) sensors offer this 
function.

i

 

Sensors of Laser Class 3B (IIIB) need an external key switch to switch off the 
laser.

  Use a serial key switch inside the control circuit to switch off the laser. You will 

find details for the wiring in the instruction manual, Chap. 5.2.4.

RS422, Synchronization

Pin no.

Assignment connector „RS422“

1

6

2

5

3

4

 

View on solder 

pin side, male 

cable connector

The pin numbers re-
fer to the gapCON-
TROL 27xx unit.

Micro-Epsilon 
recommends the 
use of the interface 
cable SC2700-4.5 
RS422 from the op-
tional accessories.

RS422

General

9-pin Sub-D

 1

1

RX1

CH1

2

2

/RX1

/CH1

3

5

GND1

5

3

TX2

CH2

9

4

/TX2

/CH2

1

6

GND2

5

The RS422 port can be used with either of the following configurations:

 - RS422: Loading of programs and transmitting of measurement results.
 - Synchronization/Triggering: Synchronization resp. triggering using switching 

signals. 

1) Optional RS422 interface cable (SC2700-4,5RS422) from Micro-Epsilon, compati-
ble to the RS422 to USB converter.  All outputs or inputs are designed symmetrically. 
Internal terminating resistors 120 Ohm.

7. 

Further Information

Please refer to the enclosed online manual. You will find details to the separate pro-
grams in the respective instruction manuals or in the instruction manual of this sensor, 
Chap. 6.2. You will find the instruction manuals online or on the provided CD.

Содержание gapCONTROL 2711

Страница 1: ...U EU directive 2011 65 EU RoHS category 9 The sensor fulfills the specifications of the EMC requirements if the instructions in the manual are followed 3 Proper Environment Protection class IP 65 Only...

Страница 2: ...required Mark the laser area recognizable and everlasting Laser Class 3B gapCONTROL 27xx sensors with a maximum laser power up to 20 mW are classified in Laser Class 3B IIIB When using lasers of 3B c...

Страница 3: ...g Data transmission Red flashing Error code The state LED indicates different error conditions by flashing If no flashing occurs for several seconds no error has occurred Power supply Power Pin no Bel...

Страница 4: ...nal accessories RS422 General 9 pin Sub D 1 1 RX1 CH1 2 2 RX1 CH1 3 5 GND1 5 3 TX2 CH2 9 4 TX2 CH2 1 6 GND2 5 The RS422 port can be used with either of the following configurations RS422 Loading of pr...

Страница 5: ...pted IP address in the link local net 169 254 x x There is no collision testing effected This is the standard setting The sensor supports the DHCP protocol This setting is activated by default and has...

Страница 6: ...System supply In 24 V 0 V OU Filter module 2 5 3 0 5 7 LINK ACT MS I O ETHERNET NS 8 7 6 5 4 3 2 1 N O LINK ACT 1 2 X 1 X 2 8 1 ON 24V X3 0V 0 WBM 255 DHCP OU Fieldbus Coupler D D Sta D D D D D L1 L2...

Страница 7: ...Connect the System supply out terminals of the OU Filter module to the System supply in terminals of the OU Fieldbus Coupler 0 V and 24 V Connect the system supply in of the OU Filter module to the p...

Страница 8: ...ct the shield of the power supply cable to the PE protective earth conduc tor of the main power supply Close unused connection plugs with the provided protective caps Switch on the power supply 24 VDC...

Отзывы: