External laser switching
Series LLT27xx(502), LLT27xx(503), LLT27xx(504) and LLT27xx(505) sensors offer this
function.
i
Sensors of Laser Class 3B (IIIB) need an external key switch to switch off the
laser.
Use a serial key switch inside the control circuit to switch off the laser. You will
find details for the wiring in the instruction manual, Chap. 5.2.4.
RS422, Synchronization
Pin no.
Assignment connector „RS422“
1
6
2
5
3
4
View on solder
pin side, male
cable connector
The pin numbers re-
fer to the gapCON-
TROL 27xx unit.
Micro-Epsilon
recommends the
use of the interface
cable SC2700-4.5
RS422 from the op-
tional accessories.
RS422
General
9-pin Sub-D
1
1
RX1
CH1
2
2
/RX1
/CH1
3
5
GND1
5
3
TX2
CH2
9
4
/TX2
/CH2
1
6
GND2
5
The RS422 port can be used with either of the following configurations:
- RS422: Loading of programs and transmitting of measurement results.
- Synchronization/Triggering: Synchronization resp. triggering using switching
signals.
1) Optional RS422 interface cable (SC2700-4,5RS422) from Micro-Epsilon, compati-
ble to the RS422 to USB converter. All outputs or inputs are designed symmetrically.
Internal terminating resistors 120 Ohm.
7.
Further Information
Please refer to the enclosed online manual. You will find details to the separate pro-
grams in the respective instruction manuals or in the instruction manual of this sensor,
Chap. 6.2. You will find the instruction manuals online or on the provided CD.