Operation
ELTROTEC Endolight FOT Halogen 150
Turn the element until the number
115
is on arrow position.
Insert the voltage selector element again and press on it.
Plug in again the power supply cable.
5.4
Fuse Replacement
Unplug the power supply cable from the socket and from the light source.
The fuse (5 x 20 mm, 2 AT) is in the voltage selector element (
8
Raise the voltage selector element, see
, at the designated cut using a screwdriver (
1
) and pull it
out (
2
).
1
2
3
Fig. 6 Fuse replacement
Remove the defective fuse (
3
) and insert a new fuse (2 AT).
Please use only the fuses recommended in the appendix under optional accessories (2 AT), see Chap.
Insert the voltage selector element again and press on it.
i
Pay attention onto the correct position of the operating voltage.
Plug in again the power supply cable.