![Michell Instruments XZR500ST Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/michell-instruments/xzr500st/xzr500st_user-manual_1784650047.webp)
XZR500 User’s Manual
Michell Instruments
39
MAINTENANCE
4.6 Error
Messages
This section explains the most common error messages that could appear during
operation:
O
2
Concentration
Storing fault / Value
1
.
This message appears when the content of the EEPROM is corrupted. This problem can
occur when the microcontroller card is replaced with a non-compatible version. All the
adjustment parameters are lost. Press RESET (
Figure 7) to cancel the message.
•
If the message does not disappear after RESET has been activated check
the microcontroller card (see Section 4.4).
Furnace temperature
Low alarm / Value
2.
The message “Low alarm” alternates every 2 seconds with the value of the temperature
measured. It will appear when the furnace temperature is 20°C below the set temperature.
Probable cause(s):
•
The analyzer was just powered up so the furnace is warming up.
Wait until the right temperature is reached.
•
There is a short-circuit between the thermocouple’s threads, causing the
analyzer to measure the temperature of the short-circuit instead of the
furnace temperature.
Check the thermocouple threads to make sure they are not short-circuiting.
•
If the red LED (heating indicator) is lit continuously, the furnace is under-
powered.
Verify the analyzer’s power supply voltage.
• Faulty
furnace.
Содержание XZR500ST
Страница 1: ...XZR500ST Oxygen Analyzer User s Manual 97137 Issue 8 April 2017...
Страница 50: ...XZR500 User s Manual 42 97137 Issue 8 April 2017 APPENDIX A Appendix A Technical Specifications...
Страница 55: ...XZR500 User s Manual Michell Instruments 47 APPENDIX C Appendix C Calculation of CO2...
Страница 59: ...XZR500 User s Manual Michell Instruments 51 APPENDIX E Appendix E Back Flushing System Optional...
Страница 62: ...XZR500 User s Manual 54 97137 Issue 8 April 2017 APPENDIX F Appendix F Heated Flue Gas Ejection System Optional...
Страница 66: ...XZR500 User s Manual 58 97137 Issue 8 April 2017 APPENDIX G Appendix G Mounting Options...
Страница 69: ...XZR500 User s Manual Michell Instruments 61 APPENDIX H Appendix H Second 4 20mA Output...
Страница 71: ...XZR500 User s Manual Michell Instruments 63 APPENDIX I Appendix I Automatic Calibration Optional...
Страница 73: ...XZR500 User s Manual Michell Instruments 65 APPENDIX J Appendix J Quality Recycling Warranty Information...
Страница 75: ...XZR500 User s Manual Michell Instruments 67 APPENDIX K Appendix K Return Document Decontamination Declaration...
Страница 77: ......
Страница 78: ...http www michell com...