XZR500 User’s Manual
Michell Instruments
37
MAINTENANCE
4.3
Replacement of the Furnace
•
Remove the probe inner tube (
Figure 4 (23)).
•
Turn off the analyzer and disconnect the power lead.
•
Wait until the furnace and the sensor housing cool down (around 45 minutes).
•
Remove the cover of the sensor housing.
•
Disconnect the cable.
•
Disconnect all the wires from the MSRS, the furnace and the connector block
(
Figure 4 (11)).
•
Remove the MSRS (
Figure 4 (5)).
•
Remove the sealing head (
Figure 4 (3)).
MSRS
O ring
Sealing
head
• Clean the sealing head if necessary.
• Loosen the probe mounting plate
fastener (
Figure 4 (19, 20)).
• Unscrew the corner plate (2 CHC
screws M6).
• Draw back the mounting plate.
• Remove the furnace.
• Place the new furnace (you may
have to enlarge the holes in the
corner plate to allow the furnace
wire’s terminals to pass).
• Replace the stop washers.
• Replace the corner plate.
•
Bring the attachment plate close to the corner plate.
•
Fasten the corner plate by pushing it down so that the lock washers lay against
the furnace upper cheek.
•
Tighten up the probe mounting plate fastener.
•
Re-insert the clean sealing head with the new MSRS (Ref 204 011) and new
o-rings (Ref 108 006)
(CAUTION: the head of the MSRS (1) has to stop
against the bearing of the sealing head body (3)). (See
Figure 16.)
•
Connect the MSRS and the cable according to the diagram
(Figure 16).
•
Replace the cover.
•
Power up the analyzer.
•
Wait for one hour.
• Re-
fi
t the probe inner tube.
•
Set self-cleaning (0.10 code).
•
Wait for a few hours before calibrating the analyzer (see Section 3.3.3).
Содержание XZR500ST
Страница 1: ...XZR500ST Oxygen Analyzer User s Manual 97137 Issue 8 April 2017...
Страница 50: ...XZR500 User s Manual 42 97137 Issue 8 April 2017 APPENDIX A Appendix A Technical Specifications...
Страница 55: ...XZR500 User s Manual Michell Instruments 47 APPENDIX C Appendix C Calculation of CO2...
Страница 59: ...XZR500 User s Manual Michell Instruments 51 APPENDIX E Appendix E Back Flushing System Optional...
Страница 62: ...XZR500 User s Manual 54 97137 Issue 8 April 2017 APPENDIX F Appendix F Heated Flue Gas Ejection System Optional...
Страница 66: ...XZR500 User s Manual 58 97137 Issue 8 April 2017 APPENDIX G Appendix G Mounting Options...
Страница 69: ...XZR500 User s Manual Michell Instruments 61 APPENDIX H Appendix H Second 4 20mA Output...
Страница 71: ...XZR500 User s Manual Michell Instruments 63 APPENDIX I Appendix I Automatic Calibration Optional...
Страница 73: ...XZR500 User s Manual Michell Instruments 65 APPENDIX J Appendix J Quality Recycling Warranty Information...
Страница 75: ...XZR500 User s Manual Michell Instruments 67 APPENDIX K Appendix K Return Document Decontamination Declaration...
Страница 77: ......
Страница 78: ...http www michell com...