1. 

OVERVIEW 

 
The PROMET EExd Sampling Systems are designed for continuous on-line measurement 
and monitoring of the moisture content of a hydrocarbon gas. The sampling systems can 
be  located  close  to  the  process  gas  sample  take-off  point  in  a  potentially  explosive 
environment designated zone 1 hazardous area. 
 
PROMET Process Moisture Analyzer Sampling Systems are specifically designed for the 
measurement of moisture content in Natural Gas. The systems are configured with the 
necessary  filtration,  pressure  letdown  and  flow  control  required  for  measurements  at 
pressure before finally venting to an atmospheric or a low pressure flare line system. A 
fast-loop bypass flow arrangement is included to reduce sample flow response time lag 
and enabling the filter to be drained automatically of any potential hydrocarbon liquids 
and hydrates formed. 
 
Two  separate  series  of  sampling  systems  are  available;  one  designed  for  the 
requirements  of  Trace  Moisture  applications  such  as  molecular  sieve  drying  at  LPG 
plants  and  one  for  the  more  general  moisture  measurements  in  Natural  Gas  from 
processing through pipeline transmission. Each of these systems can also be supplied 
as either 1 or 2 channel systems.  The more general Natural Gas systems are supplied 
with triple filtration consisting of particulate and membrane filter and glycol absorption. 
For systems intended to measure Trace Moisture levels, single filtration is provided that 
consists of a stainless steel mesh particulate filter.  
 
The  sampling  systems  gas  handling  components  are  supplied  mounted  on  a  chassis 
panel  that  can  be  fitted  within  a  stainless  steel  insulated  enclosure.  The  enclosure 
provides  environmental  protection  to  IP66  and  fitted  with  mounting  brackets

 

suitable 

for wall mounting.  
 
Sampling  systems  fitted  within  enclosures  are  temperature  controlled  to  maintain  a 
constant temperature environment of at least 10 oC above the highest envisaged dew 
point temperature independent of surrounding temperature variations. The temperature 
control system consists of a heater controlled by a fixed preset thermostat to provide 
internal  ambient  air  temperature  control  of  >15  oC.  For  higher  temperature  control, 
systems can be supplied with a heater controlled by an adjustable setpoint thermostat 
to achieve internal ambient air temperature control up to 60 oC for maintaining more 
stable measurement conditions for installations in adverse climates. 
 
All sample gas wetted metallic parts in AISI 316L stainless steel with Viton soft  parts 
that  comply  with  the  NACE  standard  MR-01-75  (latest  edition).  Tube  fittings are twin 
ferrule  compression  type.  All  gas  and  cable  entries  are  located  in  the  base  of  the 
enclosure. 
 
 
 
 
 
 

Содержание Promet EExd

Страница 1: ...2006 Michell Instruments Ltd PROMET EExd Process Moisture Analyzer Sampling Systems Installation Operation Maintenance Manual 97095 Issue 03 August 2009...

Страница 2: ...ss quality standards During development our laboratory testing team took your instrument through a rigorous testing process to ensure that it will deliver outstanding performance time and time again W...

Страница 3: ...Output Wiring Connections 12 3 OPERATION 14 3 1 Start Up Procedure 15 3 2 Shut Down Procedure 15 4 OPTIONS 16 4 1 Enclosure Heater Temperature Control 16 4 2 Trace Heated Sample Line 17 4 3 Enclosure...

Страница 4: ...ransmission Each of these systems can also be supplied as either 1 or 2 channel systems The more general Natural Gas systems are supplied with triple filtration consisting of particulate and membrane...

Страница 5: ...el BV4 for 2 channel Allows user to manually isolate the system from the process sample gas supply line for maintenance or servicing Filter s Natural Gas systems F1 for 1 channel F4 for 2 channel Prov...

Страница 6: ...re version has 2 stages of pressure letdown Trace Moisture systems PR1 PR2 for 1 channel PR3 PR4 for 2 channel Provides pressure letdown from line pressure to a lower vent pressure Standard version ha...

Страница 7: ...BV3 for 1 channel BV6 for 2 channel Allows the user to manually letdown any sample gas pressure within the system depressurise to atmosphere for maintenance or servicing Bypass flow gas handling compo...

Страница 8: ...com 1 1 Manufacturing Quality Your sampling system should reach you in perfect working condition Michell Instruments are registered with the British Standards Institute for Quality Assurance to BS EN...

Страница 9: ...for outdoor installations All are suitable for wall mounting When an outdoor system is supplied the internal insulation and temperature control circuits allow the sampling system to be positioned out...

Страница 10: ...__________________ 10 ____________________________________________________________ www michell com M8 MOUNTING HOLES 735 695 715 770 1 CHANNEL PANEL 3mm DOUBLE BIT LOCKS 1 CHANNEL ENCLOSURE M10 MOUNTI...

Страница 11: ...rate the heater circuit and analyzer within the sampling systems The factory set power supply voltage is indicated on a yellow label located on the rear panel The user cannot change the specified powe...

Страница 12: ...onto terminals within the OUTPUTS junction box Cable entry into the junction box is provided through an ISO M20 thread or EExe plastic cable gland 1 Channel Version wiring connections are as follows...

Страница 13: ...l 1 Analyzer Status Alarm Relay NO 7 Channel 2 Process Alarm Relay NC 8 Channel 2 Process Alarm Relay COM 9 Channel 2 Process Alarm Relay NO 10 Channel 2 Analyzer Status Alarm Relay NC 11 Channel 2 An...

Страница 14: ...edure stipulated as part of the hazardous area certification of the PROMET EExd product This procedure must be fully carried out prior to any power or signal connections to the PROMET EExd being energ...

Страница 15: ...Regulator PR4 for 1 channel and PR8 for 2 channel 7 For Natural Gas systems only Adjust the Bypass Flowmeter Valve FM2 for 1 channel and FM4 for 2 channel to indicate a gas flowrate of approximately 0...

Страница 16: ...ns for installations in adverse climates The capillary tube thermostat marked Enclosure Heater Thermostat contains an adjustment screw for the temperature control set point factory set to approximatel...

Страница 17: ...areas The trace heated sample line is connected directly onto the Gas Inlet Isolation Valve via bulkhead entry seal gland when fitted to an enclosure within the sampling systems and the heating cable...

Страница 18: ...ent shows that it is in poor good condition after 12 months operation then the operating period between replacements may be reduced increased accordingly To replace the membrane filter element proceed...

Страница 19: ...ndition after 12 months operation then the operating period between replacements may be reduced increased accordingly To replace the particulate filter element proceed as follows 1 Isolate the samplin...

Страница 20: ...te filter element proceed as follows 1 Isolate the sampling system from the gas supply in accordance with the System Shutdown Procedure 2 Unscrew and remove the filter bowl from the body of the filter...

Страница 21: ...ll mounting brackets Ingress protection rated to IP66 NEMA 4 Enclosure Heating Standard 100W heater with preset 20 C set point thermostat 1 heater for 1 Channel 2 heaters for 2 Channel Optional Adjust...

Страница 22: ...tenance Manual and User Guide ____________________________________________________________ 22 ____________________________________________________________ www michell com APPENDIX A CONTACT DETAILS Fo...