PROMET EExd Process Moisture Analyzer Sampling Systems
Installation, Operation & Maintenance Manual
and User
Guide
____________________________________________________________
19
____________________________________________________________
www.michell.com
1.
Isolate the sampling system from the gas supply in accordance with the System
Shutdown Procedure.
2.
Unscrew and remove the filter bowl from the body of the filter (F3).
3.
Remove and discard the black glycol filter cartridge.
4.
Refit a new glycol filter cartridge
(Michell Part No. PTS-GAC).
5.
Replace the filter bowl back into the body of the filter until it reaches the
mechanical stop.
6.
Resume normal system operation in accordance with the System Start-Up
Procedure.
5.3 Particulate Filter Element Replacement
Used on Natural Gas systems only!
Life expectancy of the filter element is dependent upon operating conditions in each
specific application. As a minimum, it is recommended that the membrane filter
element be changed every 12 months. If inspection of the removed element shows that
it is in poor/good condition after 12 months operation then the operating period
between replacements may be reduced/increased accordingly.
To replace the particulate filter element, proceed as follows:
1.
Isolate the sampling system from the gas supply in accordance with the System
Shutdown Procedure.
2.
Unscrew and remove the filter hex nut from the body of the filter (F1)
- Take
care not to lose the spring inside the nut!
3.
Carefully prise out the sintered filter element from the filter body and
clean/discard the filter element.
4.
Push fit a cleaned/new filter element
(Michell Part No. CD2-PFE).
5.
Replace the filter hex nut back onto the body of the filter until it reaches a
mechanical stop by hand – Tighten (nip) up with a spanner/wrench
- Ensure the
spring is present inside the nut!
6.
Resume normal system operation as described in the System Start-Up
Procedure.
Used on Trace Moisture systems only!
Life expectancy of the filter element is dependent upon operating conditions in each
specific application. As a minimum, it is recommended that the particulate filter element
be changed every 12 months. If inspection of the removed element shows that it is in
poor/good condition after 12 months operation then the operating period between
replacements may be reduced/increased accordingly.