3. 

OPERATION 

 
Operation of the PROMET EExd Sampling Systems should be carried out in conjunction 
with and referring to the PROMET EExd Process Moisture Analyzer Users Guide prior to 
commencing system start-up procedure. 
 
Before  commencing  the  start  up  procedure  it  is  essential  to  ensure  the  installation 
conforms to the correct Hazardous area and local plant standards. 
 
Check all cables are according to certificated approved specifications and as a minimum 
are as described below: 
 

Minimum Cable Requirements 

Power Cable 

3 core, 0.75 mm² conductor area (6A) 

Communications 
Cable 

Communications: For use with 4-20mA only or Modbus 
only - 2 pairs, 0.5 mm² conductors and with an overall 
braided screen. 

Communications 
Cable 

For use with 4-20mA and Modbus - 4 pairs individually 
screened, 0.5 mm² conductors and with a overall braided 
screen. 

 
There  is  a  mandatory  purge  procedure  stipulated  as  part  of  the  hazardous  area 
certification of the PROMET EExd product. This procedure must be fully carried out prior 
to any power or signal connections to the PROMET EExd being energised. 
 
This procedure must also be carried out at any time following service or maintenance 
periods  that  causes  any  of  the  PROMET  EExd  or  associated  gas  handling  equipment 
sample lines to be disconnected. 
 
It is not necessary to carry out this procedure if, during a shut down period, the sample 
lines have not been disconnected, or if the power or signal connections only have been 
disconnected. 
  
Before commencement of the start-up procedure ensure that all power and signal 
connections to the PROMET EExd flameproof enclosure are fully isolated and if 
necessary observe the stipulated de-energisation period of 45 minutes.   
 
Before  any  gas  pressure  is  applied,  check  all  gas  inlet  &  outlet  connections  are  fully 
tightened up and that all valves and regulators are in the closed position. 
 
BEFORE power is applied to the sampling system check that the PROMET EExd Power 
Isolator switch is set to the OFF position.  
 
Apply power to the sampling system and allow the temperature within the enclosure to 
stabilise.  This  will  prevent  the  possibility  of  condensation  occurring  within  the  gas 
handling components during start-up. 

Содержание Promet EExd

Страница 1: ...2006 Michell Instruments Ltd PROMET EExd Process Moisture Analyzer Sampling Systems Installation Operation Maintenance Manual 97095 Issue 03 August 2009...

Страница 2: ...ss quality standards During development our laboratory testing team took your instrument through a rigorous testing process to ensure that it will deliver outstanding performance time and time again W...

Страница 3: ...Output Wiring Connections 12 3 OPERATION 14 3 1 Start Up Procedure 15 3 2 Shut Down Procedure 15 4 OPTIONS 16 4 1 Enclosure Heater Temperature Control 16 4 2 Trace Heated Sample Line 17 4 3 Enclosure...

Страница 4: ...ransmission Each of these systems can also be supplied as either 1 or 2 channel systems The more general Natural Gas systems are supplied with triple filtration consisting of particulate and membrane...

Страница 5: ...el BV4 for 2 channel Allows user to manually isolate the system from the process sample gas supply line for maintenance or servicing Filter s Natural Gas systems F1 for 1 channel F4 for 2 channel Prov...

Страница 6: ...re version has 2 stages of pressure letdown Trace Moisture systems PR1 PR2 for 1 channel PR3 PR4 for 2 channel Provides pressure letdown from line pressure to a lower vent pressure Standard version ha...

Страница 7: ...BV3 for 1 channel BV6 for 2 channel Allows the user to manually letdown any sample gas pressure within the system depressurise to atmosphere for maintenance or servicing Bypass flow gas handling compo...

Страница 8: ...com 1 1 Manufacturing Quality Your sampling system should reach you in perfect working condition Michell Instruments are registered with the British Standards Institute for Quality Assurance to BS EN...

Страница 9: ...for outdoor installations All are suitable for wall mounting When an outdoor system is supplied the internal insulation and temperature control circuits allow the sampling system to be positioned out...

Страница 10: ...__________________ 10 ____________________________________________________________ www michell com M8 MOUNTING HOLES 735 695 715 770 1 CHANNEL PANEL 3mm DOUBLE BIT LOCKS 1 CHANNEL ENCLOSURE M10 MOUNTI...

Страница 11: ...rate the heater circuit and analyzer within the sampling systems The factory set power supply voltage is indicated on a yellow label located on the rear panel The user cannot change the specified powe...

Страница 12: ...onto terminals within the OUTPUTS junction box Cable entry into the junction box is provided through an ISO M20 thread or EExe plastic cable gland 1 Channel Version wiring connections are as follows...

Страница 13: ...l 1 Analyzer Status Alarm Relay NO 7 Channel 2 Process Alarm Relay NC 8 Channel 2 Process Alarm Relay COM 9 Channel 2 Process Alarm Relay NO 10 Channel 2 Analyzer Status Alarm Relay NC 11 Channel 2 An...

Страница 14: ...edure stipulated as part of the hazardous area certification of the PROMET EExd product This procedure must be fully carried out prior to any power or signal connections to the PROMET EExd being energ...

Страница 15: ...Regulator PR4 for 1 channel and PR8 for 2 channel 7 For Natural Gas systems only Adjust the Bypass Flowmeter Valve FM2 for 1 channel and FM4 for 2 channel to indicate a gas flowrate of approximately 0...

Страница 16: ...ns for installations in adverse climates The capillary tube thermostat marked Enclosure Heater Thermostat contains an adjustment screw for the temperature control set point factory set to approximatel...

Страница 17: ...areas The trace heated sample line is connected directly onto the Gas Inlet Isolation Valve via bulkhead entry seal gland when fitted to an enclosure within the sampling systems and the heating cable...

Страница 18: ...ent shows that it is in poor good condition after 12 months operation then the operating period between replacements may be reduced increased accordingly To replace the membrane filter element proceed...

Страница 19: ...ndition after 12 months operation then the operating period between replacements may be reduced increased accordingly To replace the particulate filter element proceed as follows 1 Isolate the samplin...

Страница 20: ...te filter element proceed as follows 1 Isolate the sampling system from the gas supply in accordance with the System Shutdown Procedure 2 Unscrew and remove the filter bowl from the body of the filter...

Страница 21: ...ll mounting brackets Ingress protection rated to IP66 NEMA 4 Enclosure Heating Standard 100W heater with preset 20 C set point thermostat 1 heater for 1 Channel 2 heaters for 2 Channel Optional Adjust...

Страница 22: ...tenance Manual and User Guide ____________________________________________________________ 22 ____________________________________________________________ www michell com APPENDIX A CONTACT DETAILS Fo...

Отзывы: