background image

PC33 & 52 User’s Manual

18

  

 

97238 Issue 2, October 2013

appendix d

• 

Cleanliness  of  the  environment  will  affect  different  hygrometers  in  different 

ways. Dust and airborne droplets should be avoided or filtered out if possible. 

Contaminants  can  come  from  the  most  surprising  sources,  ordinary  urban 

pollution, for example.

• 

The readings given by some types of hygrometer are sensitive to gas type. For 

any Instrument which reads in terms of mass per unit volume, e.g. in grams per 

cubic metre, it must be confirmed whether the calibration is valid for the gas in 

use.

• 

Avoid  using  instruments  in  direct  sunlight  or  near  any  other  source  of  heat, 

unless they are suitably shielded to prevent measurement errors.

Sampling in general

• 

Relative  humidity  measurements  should  be  carried  out  at  a  representative 

temperature.  Failure  to  allow  temperature  equilibration  will  lead  to  a  false 

indication of the relative humidity.

• 

Variations in vapor pressure from place to place can occur where an environment is 

subject to any addition or removal of water. If so, care must be taken over where 

to make a measurement in order to obtain a representative result.

• 

Sources  and  sinks  of  water  vapor  should  be  avoided  in  any  sampling  system. 

Invasion  of  stray  water  can  be  minimised  by  attention  to  leaks,  hygroscopic 

materials, droplets and condensation. The lower the humidity, the more critical 

these precautions are.

• 

Hygroscopic materials should be avoided. Many materials contain moisture as part 

of their structure, particularly organic materials (whether natural or synthetic), 

salts (or anything which contains them), and anything which has small pores. 

Temperature changes can increase the tendency of these materials to affect the 

humidity of the surrounding air.

• 

Condensation in a sampling process can invalidate humidity measurements by 

reducing the water content of the gas being measured. What is more, condensed 

liquid may alter the humidity elsewhere by dripping or running to other locations 

and  evaporating  there.  In  these  circumstances,  measurement  results  may  be 

misleading if hygrometer location is not considered carefully.

• 

Water droplets or mist must be avoided. These can result in overestimates of the 

humidity of the air between the droplets. Such results may exceed 100% RH, 

or may be impossible to interpret meaningfully. Droplets of liquid also damage 

some electrical types of humidity sensor. Filtering the air sample can eliminate 

droplets.

• 

If pumps are used for sampling gas, these should be located after the hygrometer, 

to  avoid  contaminating  the  measurement  environment.  Where  possible,  oil 

free  pumps  should  be  used,  or  filters  employed.  Oscillations  in  pressure  due 

to  pumping  can  sometimes  be  reduced  or  buffered  using  a  needle  valve  or  a 

reservoir of large volume.

• 

Special treatments such as filtration can change the amount of moisture in a gas. 

Some drying agents take out other gases too.

• 

When sealing any sensor or probe into a port or manifold in a duct or chamber, 

leaks through the probe or electrical cable should be considered. These are not 

always sealed against passage of ambient air.

• 

Where  sampling  involves  a  step  change  in  temperature,  pressure  or  gas  flow 

rate, relative to the process being sampled, results may need to be converted 

or interpreted. For example ‘pressure dew point’ will differ from the value found 

after expanding the gas sample to atmospheric pressure. Care should be taken 

to distinguish between ‘gauge’ and absolute values of pressure.

Содержание PC33

Страница 1: ...PC33 52 RH Temperature Probes User s Manual M I C H E L L IN S T R U M E N T S E L I T R O N I C L I W 4 97238 Issue 2 October 2013...

Страница 2: ...nformation when contacting Michell Instruments for service purposes Instrument Code Serial Number Invoice Date Location of Instrument Tag No Instrument Code Serial Number Invoice Date Location of Inst...

Страница 3: ...property of Michell Instruments Ltd and may not be copied or otherwise reproduced communicated in any way to third parties nor stored in any Data Processing System without the express written authori...

Страница 4: ...ent 6 Figures Figure 1 PC33 52 1 Figure 2 Cable Connections 2 Figure 3 Adjustment Tube 3 Figure 4 PC33 2 wire Connection 4 Figure 5 PC33 3 4 wire Connection 5 Figure 6 PC52 2 wire Connection 6 Figure...

Страница 5: ...for the user to come into contact with any hazardous substance which might be employed in the construction of the instrument Care should however be exercised during maintenance and the disposal of ce...

Страница 6: ...m millimeters oz ounce s RH relative humidity T temperature V Volts Ohms inch es percentage Warnings The following general warning listed below is applicable to this instrument It is repeated in the t...

Страница 7: ...pplications Figure 1 PC33 52 1 1 Highlights Low cost PC33 with analog output is designed for HVAC applications PC52 with analog output is designed for accurate measurement in controlled environments A...

Страница 8: ...tor White Pin 1 Power Supply V Yellow Pin 2 Output temperature Brown Pin 3 Common ground Green Pin 4 Output RH 4 20 mA output 2 wire Cable Connector White Pin 1 Output RH Yellow Pin 2 Output temperatu...

Страница 9: ...nd high temperature conditions more frequent calibration is recommended If re adjustment is necessary refer to Sections 3 1 and 3 2 and also see the drawings of the circuit boards for the location of...

Страница 10: ...ached and the reading of the reference has stabilized adjust the transmitter with the RH potentiometer see 1 below For versions with a T output the temperature should be calibrated at ambient temperat...

Страница 11: ...TOR CABLE RH Temp REL HUMIDITY OUTPUT J5 J6 0 to 1 V DC 0 to 5 V DC 0 to 10 V DC Soldered joint TEMPERATURE RANGE OUTPUT 0 to 50 C 32 to 122 F 0 to 100 C 32 to 212 F 20 to 80 C 4 to 176 F Output J7 J8...

Страница 12: ...e transmitter with the Zero RH potentiometer see 1 below Compare with the reference After the second high value is reached and stabilized adjust with the Span RH potentiometer see 2 below For versions...

Страница 13: ...CABLE RH Temp REL HUMIDITY OUTPUT J1 J2 J3 0 to 1 V DC 0 to 5 V DC 0 to 10 V DC Soldered joint SETTINGS TEMPERATURE OUTPUT AND RANGE WITH SOLDER JUMPERS 0 to 50 C 32 to 122 F 0 to 100 C 32 to 212 F 2...

Страница 14: ...PC33 52 User s Manual 8 97238 Issue 2 October 2013 appendix a Appendix A Technical Specifications...

Страница 15: ...Electrical Specifications Output Signal Options 4 20 mA 0 to 1 0 to 5 0 to 10 V Supply Voltage 14 to 30 V DC for 0 to 5 0 to 10 V output 5 to 30 V DC 0 to 1 V mA output Operating Specifications Operat...

Страница 16: ...PC33 52 User s Manual 10 97238 Issue 2 October 2013 appendix B Appendix B EC Declaration...

Страница 17: ...directives listed below 2004 108 EC EMC Directive Using the standards EN61326 1 2006 Electrical equipment for measurement control and laboratory use EMC requirements Group 1 Class B emissions and imm...

Страница 18: ...s listed below 2004 108 EC EMC Directive Using the standards EN61326 1 2006 Electrical equipment for measurement control and laboratory use EMC requirements Group 1 Class B emissions and immunity and...

Страница 19: ...PC33 52 User s Manual Michell Instruments 13 appendix C Appendix C Quality Recycling Warranty Information...

Страница 20: ...tive and the Restriction of Hazardous Substances RoHS Directive place rules upon European manufacturers of electrical and electronic equipment The directives aim is to reduce the impact that electroni...

Страница 21: ...escription of the fault and the customer contact information 3 Upon receipt Michell Instruments will evaluate the product to determine the cause of the malfunction Then one of the following courses of...

Страница 22: ...PC33 52 User s Manual 16 97238 Issue 2 October 2013 appendix D Appendix D Recommended Practices in Humidity Measurement...

Страница 23: ...instrument for the job Allow hygrometers to equilibrate in any new environment This is particularly necessary after changes in temperature due to transportation or storage Depending on the instrument...

Страница 24: ...ature changes can increase the tendency of these materials to affect the humidity of the surrounding air Condensation in a sampling process can invalidate humidity measurements by reducing the water c...

Страница 25: ...ments Ambient humidity Avoid using hygrometers near the body which is a source of heat and moisture Do not breathe close to the measurement High humidity above the ambient range sample lines should be...

Страница 26: ...types of hygrometer Relative humidity capacitive sensor Care should be taken to avoid mechanical shock impact or thermal shock sudden temperature changes Sensors should be protected from steam or wate...

Страница 27: ...PC33 52 User s Manual Michell Instruments 21 NOTES...

Страница 28: ...http www michell com...

Отзывы: