PC33 & 52 User’s Manual
4
97238 Issue 2, October 2013
CALIBRATION
3.1
Calibration & Adjustment for 3% Accuracy (1pt Adjustment)
PC33
1.
Remove the screw at the side of the connector.
2.
Carefully unscrew the sensor protection cap from the head.
dO nOT TOUCH THE SEnSOR ELEMEnT
(see 3 below)
3.
Carefully remove the circuit board from the body by pulling on the
threaded end while holding the tube.
•
The transmitter should be calibrated at one point, for example 50%
RH.
•
Once this value is reached and the reading of the reference has
stabilized, adjust the transmitter with the RH potentiometer (see
1 below).
•
For versions with a T output the temperature should be calibrated
at ambient temperature.
•
After stabilizing, adjust with the temperature potentiometer (see
2 below).
See
Figure 5 for 2-wire V DC power supply voltage, output 4-20 mA and Figure 6 for
3/4-wire V DC power supply voltage, output 0 to 1, 0 to 5 or 0 to 10 V.
2-wire (mA)
1
3
2
PIN 1
WHITE
PIN 3
BROWN
PIN 4
GREEN
PIN 2
YELLOW
CONNECTIONS
RH
-
T
-
CONNECTOR
CABLE
TEMPERATURE RAnGE
INPUT
j3
j4
j5
0 to +50°C
(+32 to +122°F)
0 to +100°C
(+32 to +212°F)
-20 to +80°C
(-4 to +176°F)
Soldered joint
Figure 4
PC33 - 2-wire Connection
Содержание PC33
Страница 14: ...PC33 52 User s Manual 8 97238 Issue 2 October 2013 appendix a Appendix A Technical Specifications...
Страница 16: ...PC33 52 User s Manual 10 97238 Issue 2 October 2013 appendix B Appendix B EC Declaration...
Страница 19: ...PC33 52 User s Manual Michell Instruments 13 appendix C Appendix C Quality Recycling Warranty Information...
Страница 27: ...PC33 52 User s Manual Michell Instruments 21 NOTES...
Страница 28: ...http www michell com...