Optidew Series User’s Manual
24
97430 Issue 2, March 2017
OPERATION
The Calibration Correction
window is a utility that allows an authorized user to input
calibration information in order to effect a real-time correction of the displayed, charted
and logged data within the Opti-Soft application software.
Data for dew-point temperature and ambient temperature, both in units of °C, may be
entered for correction purposes, along with the original Calibration Certi
fi
cate reference
number and date of calibration, providing full traceability of data. Once the correction
data has been applied, by clicking on the check box, the main Virtual Hygrometer
window will indicate that corrected data is being displayed and will show the Calibration
Certi
fi
cate number and date. This information is also saved to the
Log
fi
le for data
export.
Figure 11 shows the Calibration Correction window. Four sets of data may be entered:
DP Ref
Dew-point data for the reference hygrometer (sometimes
called the actual dew point or the standard)
DP Reading
Measured dew-point value of the Optidew under test
Temp Ref
Temperature data from the reference thermometer
Temp Reading
Measured temperature value of the Optidew under test
Data can be entered for between 3 and 11 different dew-point and temperature
calibration points. If no data is inserted, no calibration correction is possible. Data
should be entered with the highest dew-point and temperature values at the top of
the page, in descending value order to the bottom. If out-of-sequence data or spurious
characters are entered, the software will raise a warning message and bad data must
be re-entered.
The
Calibration Certificate Number
section is an optional entry field and is
alphanumeric. Any information entered into this field will be displayed on the main
Virtual Hygrometer window when calibration correction is enabled. In addition it will
be saved to the
Log
file. Similarly, the date of calibration may be entered for display
and logging when correction is enabled.
Содержание Optidew
Страница 41: ...Optidew Series User s Manual Michell Instruments 33 MAINTENANCE Appendix A Technical Specifications ...
Страница 44: ...Optidew Series User s Manual 36 97430 Issue 2 March 2017 APPENDIX A Appendix B Optidew RS232 Commands ...
Страница 48: ...Optidew Series User s Manual 40 97430 Issue 2 March 2017 APPENDIX B Appendix C Troubleshooting Common Faults ...
Страница 54: ...Optidew Series User s Manual 46 97430 Issue 2 March 2017 APPENDIX D Appendix D Dimensional Drawings ...
Страница 63: ...NOTES ...
Страница 64: ......
Страница 65: ......
Страница 66: ...http www michell com ...