Optidew Series User’s Manual
Michell Instruments
17
OPERATION
3.3.7 Mirror Condition and Peltier Power
Peltier Power indicates how much the heat pump is depressing in order to measure
the dew point. When the peltier power has a value of 100% and does not reduce over
an extended period of time, it means that the heat pump is at maximum depression.
In normal operation this indicates that the dew point is lower than the present
mirror temperature and therefore cannot be measured. Reducing the sensor ambient
temperature by use of additional cooling will increase the measurement range of the
instrument in applications where the peltier power >95%.
NOTE: Operating the sensor at the limits of its depression capability for
extended periods may cause premature wear to the heat pump.
Alongside the peltier power value is an indicator that shows the control stability. When
this indicator shows CNTRL, it indicates the system is controlling the mirror temperature
on the dew point. COOL indicates the system is depressing the heat pump in order to
form dew on the mirror surface. HEAT indicates a rapid increase in dew-point level,
whereby the system needs to increase the temperature of the mirror surface to read
this new dew-point value.
The Mirror Condition indicates the amount of signal received back from the mirror which
includes both the level of moisture and contamination on the mirror surface. In DCC
mode this display will only show the amount of mirror contamination and, if greater
than 80% after a DCC, will initiate an optics alarm condition.
NOTE: It is recommended to clean the mirror and reset the mirror condition
before an Optics fault occurs. Refer to Section 5 for further instructions.
3.4
Optisoft Application Software
The Opti-Soft application software is an interface to the Optidew Series that provides
a display of the measured and calculated parameters, system status, charting and
logging, statistical information and a facility to view and change the system parameters.
NOTE: If your Optidew Series instrument is fitted with a display,
Communication with the software is only possible when the unit is in
REMOTE
MODE
(see Section 3.2).
Содержание Optidew
Страница 41: ...Optidew Series User s Manual Michell Instruments 33 MAINTENANCE Appendix A Technical Specifications ...
Страница 44: ...Optidew Series User s Manual 36 97430 Issue 2 March 2017 APPENDIX A Appendix B Optidew RS232 Commands ...
Страница 48: ...Optidew Series User s Manual 40 97430 Issue 2 March 2017 APPENDIX B Appendix C Troubleshooting Common Faults ...
Страница 54: ...Optidew Series User s Manual 46 97430 Issue 2 March 2017 APPENDIX D Appendix D Dimensional Drawings ...
Страница 63: ...NOTES ...
Страница 64: ......
Страница 65: ......
Страница 66: ...http www michell com ...