Liquidew I.S. User Manual
Michell Instruments
23
INSTALLATION
2.4.3.2 Temperature Transmitter Wiring
HAZARDOUS AREA INFORMATION
The temperature transmitter is certified intrinsically safe for
use in Hazardous areas.
Before using the temperature transmitter in any hazardous
environment, ensure that personnel are completely familiar
with the above standards relating to the certification of this
instrument and also with the intrinsically safe apparatus
information in EN60079-14:1997 or equivalent codes of
practice in the country of installation.
If the temperature transmitter is not ordered together with the
analyzer, it is the user’s responsibility to make sure that the
temperature transmitter is compatible with the I.S. barrier in
the control unit.
Refer to ATEX/UKCA/QPS/IECEx certificates for the dew-point
and temperature transmitters’ connection cable requirements
which stipulate maximum permissible mutual capacitance and
inductance to resistance ratio.
a. Cable connection to the temperature transmitter is made via the terminal
block (4) accessed by removing the terminal housing lid (2)
(Figure 17).
1
1
2
3
4
5
Figure 17
Temperature Transmitter Housing Example
b.
Ensure that the outer diameter of the selected cable is matched to an
EExe M20 cable gland (5). Unscrew the cable gland (5) and slide the cable
through the cable gland (5) and into the terminal housing (1) through the
cable entry (3).
Содержание Liquidew I.S.
Страница 1: ...97092 Issue 6 July 2022 Liquidew I S Process Moisture Analyzer User Manual...
Страница 41: ...Liquidew I S User Manual Michell Instruments 33 APPENDIX A Appendix A Technical Specifications...
Страница 44: ...Liquidew I S User Manual 36 97092 Issue 6 July 2022 APPENDIX B Appendix B Serial Communications...
Страница 55: ...Liquidew I S User Manual Michell Instruments 47 APPENDIX C Appendix C Hazardous Area Certification...
Страница 58: ...Liquidew I S User Manual 50 97092 Issue 6 July 2022 APPENDIX D Appendix D System Drawings...
Страница 61: ...Liquidew I S User Manual Michell Instruments 53 APPENDIX E Appendix E Quality Recycling Warranty Information...
Страница 65: ...Liquidew I S User Manual Michell Instruments 57 NOTES...
Страница 66: ...http www michell com www ProcessSensing com...