Liquidew I.S. User Manual
Michell Instruments
21
INSTALLATION
Preparation of the Sensor Cable
In order to comply with hazardous area certification of the
product it is essential that the crimps/wire ferrules supplied
must be attached on to any cable installed into the connector.
a. As shown in
Figure 13 below, the crimps/wire ferrules should be applied
so that there is no possibility of a conductor strand of a core becoming
free.
When the crimp/wire ferrules are applied they should have a minimum of
two positions of crimping. After the crimp/wire ferrules are applied they
should be trimmed to a length of 5 mm (0.2”) (see
Figure 14).
10
mm
Figure 13
Crimped Wires
Figure 14
Cut to 5 mm
b. Cable connection to the dew-point transmitter is made via the terminal
block (4) (see
Figure 15). Remove the terminal housing lid (2) to access.
1
4
5
1
3
2
Figure 15
Dew-Point Transmitter Housing
Содержание Liquidew I.S.
Страница 1: ...97092 Issue 6 July 2022 Liquidew I S Process Moisture Analyzer User Manual...
Страница 41: ...Liquidew I S User Manual Michell Instruments 33 APPENDIX A Appendix A Technical Specifications...
Страница 44: ...Liquidew I S User Manual 36 97092 Issue 6 July 2022 APPENDIX B Appendix B Serial Communications...
Страница 55: ...Liquidew I S User Manual Michell Instruments 47 APPENDIX C Appendix C Hazardous Area Certification...
Страница 58: ...Liquidew I S User Manual 50 97092 Issue 6 July 2022 APPENDIX D Appendix D System Drawings...
Страница 61: ...Liquidew I S User Manual Michell Instruments 53 APPENDIX E Appendix E Quality Recycling Warranty Information...
Страница 65: ...Liquidew I S User Manual Michell Instruments 57 NOTES...
Страница 66: ...http www michell com www ProcessSensing com...