HygroCal100 User’s Manual
Michell Instruments
33
MAINTENANCE
4.4 Calibration
Correction
The HygroCal100 is provided with an intuitive calibration correction system which is used
to calibrate and adjust the internal HS3 control sensor against an external reference
device.
If a traceable reference is available, it is advisable to pass the traceability of this onto
the control sensor of the HygroCal100 at regular intervals. The process allows calibration
at 3, 5 or 9 preset points, which are:
Preset
RH Points
Total time to
complete
3 point
30%, 50%, 70%
45mins
5 point
20%, 30%, 50%, 70%, 80%
1hr 15mins
9 point
10%, 20%, 30%, 40%, 50%, 60%, 70%, 80%, 90%
2hr 15mins
Table 11
Calibration Correction
Select System Settings ->
Correct.
This will display a prompt to enter a password
to proceed - the required code is
7316
. Once this has been entered, the Calibration
Correction Setup Screen will be displayed. This screen shows a number of checks which
should be carried out before proceeding.
NOTE: If the desiccant is exhausted, or the saturator is not suffi ciently fi lled,
then this can result in an invalid correction being applied to the control sensor.
NOTE: Ensure that the temperature of the environment you intend to operate
in is stable before proceeding.
13:04
Calibration Correction
- Check saturator
- Check desiccant
- Is external reference installed?
Calibration Points
TRANSFER
TEMP
3
DEFAULT
START
Figure 33
Calibration Correction Setup Screen
The Calibration Correction Setup Screen allows selection of the desired number
of calibration points. Once this choice is made, touch the
Start
button to begin the
calibration.
Previous calibration corrections can be downloaded to USB memory by touching the
Transfer
icon, and will be stored in CSV format as per the log data fi les (see Section
3.9.2 for further information).
Содержание HygroCal100
Страница 1: ...HygroCal100 Humidity Validator User s Manual 97500 Issue 3 January 2018 ...
Страница 44: ...HygroCal100 User s Manual 36 97500 Issue 3 January 2018 APPENDIX A Appendix A Technical Specifications ...
Страница 46: ...HygroCal100 User s Manual 38 97500 Issue 3 January 2018 APPENDIX B Appendix B Troubleshooting ...
Страница 52: ...HygroCal100 User s Manual 44 97500 Issue 3 January 2018 NOTES ...
Страница 54: ...http www michell com ...