HygroCal100 User’s Manual
Michell Instruments
23
OPERATION
3.4.5.3 About
10:55
About
Firmware Version
© Copyright Michell Instruments B.V.
1.1.4
Last Calibration Date
24-02-2015
Figure 24
About Screen
Displays Firmware version, and date of last internal calibration correction.
3.4.5.4 Correct
13:04
Calibration Correction
- Check saturator
- Check desiccant
- Is external reference installed?
Calibration Points
TRANSFER
TEMP
APPLY
DEFAULT
Figure 25
Calibration Correction Screen
In order to enter the Calibration Correction Screen, a prompt box will appear asking the
user to enter a code. To proceed, enter
7316
.
The Calibration Correction Screen allows the initialization of an internal calibration
routine. The number of calibration points can be changed between 3, 5 and 9.
The data collected from previous calibrations can be transferred to installed USB memory
by means of the
Transfer
button.
The calibration correction can be reset to the default factory values, by means of the
Default
button.