I.S. Field Display User’s Manual
28
97321 Issue 2.3, May 2016
MAINTENANCE
4 MAINTENANCE
4.1
Fault Finding During Commissioning
If the I.S. Field Display fails to function during commissioning, the following procedure
should be followed:
Symptom
Cause
Solution
No display
Incorrect wiring
Check wiring
There should be 0.6 to 1.2 V between terminals 1 &
3 with terminal 1 positive
With an optional loop-powered backlight, there
should be 3.4 to 5 V between terminals 1 & 13 with
terminal 1 positive
No display 0 V between
terminals 1 & 3
Incorrect wiring or no
power supply
Check supply voltage and voltage drop caused by all
the instruments in the loop
All decimal points
fl
ashing Under range if –ve sign
displayed or over range
Recalibrate the numerical display
Unstable display
4-20 mA input is noisy
Eliminate ripple on 4-20 mA power supply and/or
decrease resolution
Unable to enter
Con
fi
guration Menu
Incorrect security code
entered
Enter correct security code, or contact Michell
Instruments if the code has been lost
4.2
Fault Finding After Commissioning
!
Ensure plant safety before starting maintenance.
Live maintenance is permitted on intrinsically safe equipment installed in a gas hazardous
area, but only certi
fi
ed test equipment should be used unless a gas clearance certi
fi
cate
is available.
If an I.S. Field Display fails after it has been functioning correctly follow the procedure
shown in Section 4.1. If this does not reveal the cause of the fault, it is recommended
that the instrument is replaced. This can be done without disconnecting power, but
while the I.S. Field Display is disconnected the 4-20 mA loop will be open circuit.
4.3 Servicing
All I.S. Field Displays are interchangeable if the required optional backlight and
alarms are
fi
tted. A single spare instrument may quickly be recalibrated to replace any
instrument that is damaged or fails. No attempt should be made to repair instruments
at component level.
It is recommended that faulty instruments are returned to Michell Instruments or to a
local Michell Instrument's agent for repair.
4.4 Routine
Maintenance
The mechanical condition of the instrument and electrical calibration should be regularly
checked. The interval between inspections depends upon environmental conditions. It
is recommended that, initially, instrument calibration should be checked annually.
Содержание Easidew PRO
Страница 16: ...I S Field Display User s Manual 10 97321 Issue 2 3 May 2016 OPERATION Figure 7 Configuration Menu...
Страница 35: ...I S Field Display User s Manual Michell Instruments 29 APPENDIX A Appendix A Hazardous Area Certification...
Страница 38: ...I S Field Display User s Manual 32 97321 Issue 2 3 May 2016 APPENDIX A...
Страница 39: ...I S Field Display User s Manual Michell Instruments 33 APPENDIX A...
Страница 40: ...I S Field Display User s Manual 34 97321 Issue 2 3 May 2016 APPENDIX A...
Страница 41: ...I S Field Display User s Manual Michell Instruments 35 APPENDIX A...
Страница 42: ...I S Field Display User s Manual 36 97321 Issue 2 3 May 2016 APPENDIX A...
Страница 43: ...I S Field Display User s Manual Michell Instruments 37 APPENDIX A...
Страница 44: ...I S Field Display User s Manual 38 97321 Issue 2 3 May 2016 APPENDIX A...
Страница 45: ...I S Field Display User s Manual Michell Instruments 39 APPENDIX A...
Страница 46: ...I S Field Display User s Manual 40 97321 Issue 2 3 May 2016 APPENDIX A...
Страница 47: ...I S Field Display User s Manual Michell Instruments 41 APPENDIX A...
Страница 48: ...I S Field Display User s Manual 42 97321 Issue 2 3 May 2016 APPENDIX B Appendix B EC Declaration of Conformity...
Страница 57: ...I S Field Display User s Manual Michell Instruments 51 NOTES...
Страница 58: ...http www michell com...