Condumax II Transportable
Sampling System User’s Manual
8
97438 Issue 3, April 2019
OPERATION
3.1
Sample Flow Start Up Procedure
1. Unscrew and remove the Condumax II flameproof enclosure cover after
first loosening the hex grub screw.
2. The user interface/display assembly uses two ¼-turn bayonet style
fasteners, (finger operated), to secure it - turn clockwise to lock, counter-
clockwise to release. The display must be supported once it is removed
from the fasteners. The right hand fastener can be positioned to the left
hand mount, as shown below.
Figure 6
Removal of Interface/Display
3. Locate the solenoid valve and the manual override operating adjuster
(brass screw handle) mounted on the base of the solenoid valve body.
Figure 7
Solenoid Valve Location