41
ERROR INDICATION
An error has occurred when the LED (2) of the battery gauge
blinks after switching on the wheelchair.
The LEDs (1)+(3) are lit in red, the number of blinking im-
pulses of LED (2) indicates an existing error according to the
following table Error indication.
Table Error indication
Number of
blink impuls-
es
Cause
Required measure
1
–
–
2
Left Hall sensor
Conduct a calibration, if required send in / replace the ac-
companying person control.
3
Right Hall sensor
Conduct a calibration, if required send in / replace the ac-
companying person control.
4
Left gas lever
Conduct a calibration, if required send in / replace the ac-
companying person control.
5
Right gas lever
Conduct a calibration, if required send in / replace the ac-
companying person control.
6
Faulty initialisation
Repeat diagnostic test.
7
Faulty CPU
Send in / replace accompanying person control.
8
Faulty source code
Send in / replace accompanying person control.
9
Software watchdog
Send in / replace accompanying person control.
10
R-NET internal error
Read out R-NET error protocol.
11
Diagnostic test failed
Repeat diagnostic test.
Conduct calibration
1. Disconnect R-NET adjustment module from the system!
2. Depress S2, S3 and S4 simultaneously and then press S1.
3. Keep hold of S2, S3 and S4 until a beep sound is heard
(about 5 sec.).
4. LED 1 is lit -> move the right gas lever completely once.
5. LED 2 is lit -> move the left gas lever completely once.
6. LED 3 is lit -> move the steering column completely to
the left once.
7. LED 4 is lit -> move the steering column completely to
the right once.
8. Calibration is completed.