-12-
instrument.
The conductor to be tested should be in the center position of the
flexible coil (near point A), not near the opening and closing
mouth (C area). The test error at the opening and closing mouth is
about doubled or more, and the position error near point B is
increased by about 1%.
3)
When the measured current is small, the test resolution can be
improved by repeatedly wrapping the flexible coil on the wire to be
tested and then closing it a plurality of turns. For example, the
measured current is 0.01A. After the flexible coil wraps around the
tested conductor for 3 turns, the test meter displays 0.03A, and
around 5 turns, it displays 0.05A, and so on. The actual current is
equal to the meter display value divided by the flexible coil. The
number of turns around. See the legend for the reference.
Detecting the
total electricity
safety status
Multi-turn
test
Grounding
Detecting hidden
dangers of buried
wall electricity
charges
Test
equipment
operation safety
status