The current of the line is measured by clamping the live
wire or the neutral wire separately (note that clamping 1
wire).
Clamp the live & neutral wires together to measure the
leakage current of the single phase (note that clamping 2
wires).
Clamp the ground wire to measure the leakage current of
the grounding wire of the electrical equipment (note that
clamping 1 wire).
Clamp the three-phase 3-wire together to measure the
leakage current of three-phase 3-wire (note that clamping 3
wires).
Clamp the three-phase 4-wire together to measure the
leakage current of three-phase four-wire (note that
clamping 4 wires).
Clamp the three-phase 5-wire together to measure the
leakage current of three-phase 5-wire (note that clamping 5
wires).
For safety, when measuring high voltage and high current,
after confirming that the test has been performed correctly,
move the meter away from the conductor under test.
9
.
Real Time Monitoring
Open the host to enter the test state, connect the computer and the host
computer with a randomly configured USB communication cable, and run
the monitoring software installed in the computer. If the communication is
normal, the computer can monitor the online current in real time. If the
user fails to connect to the monitoring software, there may be a lack of
driver. Install the USB to serial port driver in the randomly configured CD
to use it normally.