SERVICING
Unpacking
The complete unit content:
•
ST88 NEO
•
Bath and/or globule receptacle(s)
•
Power cord
•
Serial cord
•
Pot + ring + alloy
•
4 tool holders (D19, D23, D10 and D10D)
•
45° tool angle.
•
4 allen keys
•
One calliper
•
One bubble level
•
Tweezer
•
This manual
•
Calibration certificate
•
Software on CD
•
Video cam (optional)
Installation
The startup is done under the METRONELEC's responsibility or one of the authorized agent's.
Work station environment
To avoid the disturbing of the signal during the measurement, the apparatus shall be on a
stable working bench top, free of vibrations, away from air draughts and noises.
The room must be free of acid or corrosive vapors.
Main supply
•
230 V 50 Hz (or 110 V 50-60 Hz optional), stabilized non trouble.
•
Power 700 W
•
Inert gas supply if requested.
Powering on the unit
In fact, you have to connect your USB cable to the left plug looking in this way.
Notice the 2 other plugs have been kept as spares if needed, but they will not be used.
The right plug is notified with a sticker.