Measuring the Gamma LV
The parameters definition is similar to other applications, except for the types of sample,
only type 1 is available, as this requires the use of very thin copper foils of 10 mm width.
The data files are different than other test files.
This application is really useful to determine unknown GammaLV values.
The result is expressed in mN/m, giving the max value for the current curve (surf.tens. max)
and the max value of all values (Surf.Tens. max max).
Measuring fluxes & alloys efficiency
The parameters definition is similar to other applications, except for the types of sample only
type 1 is available, as this requires the use of very thin copper foils of 10 mm width.
The data files are different than other test files.
For this
application, we have one more field “grade” where we defined the grade / ageing of
the copper foil used. The ageing is defined in standard NFC 90 550 and NFC 90 551.
We can measure the flux spectrum of activity vs aged copper grade.
The results are expressed in mN; Mil, IEC and IPC standard are not available.