The menu Parameters Measure in progress will give you the conditions of a previous test
Keep in mind that dimensions, density and Gamma LV values are used to calculate the pass
fail limits, any error in it will create false limits. (See equations in section 1.1)
TYPE OF MEASURMENT
Choose the bath or one globule, some options are available / limited or not following this
choice. Check the receptacle chosen is well installed on the ST88 NEO !
To change the receptacle, pull up the manual jig on the right front side to free it, lift up
gently until you can move it outside the machine. Put the new receptacle by the same way,
be careful to avoid any chock with the height gauge.
The globule receptacle has three globules of 4, 2 and 1 mm diameter, the operator choose
by alignment of the desired globule with the component.
This receptacle has also a green button to confirm the alignment with one globule,
mandatory to allow the system to continue the test. This is the moment when the Motor
Control software will start. if you have putted 1 in the last lane of the menis.ini you should
have this black window with several lines, (more than 2, coming), if only two lines it means
that you have another small window behind this black one saying that pipe was not possible.
In this case you have to select in the main software the action : Measure / relancer le
module MC. Please see the two following windows :