7 Technical specifications
■■■■■■■■
37
Wavelength accu-
racy
< 0.05 nm (NIST SRM 1920)
Wavelength preci-
sion
Based on one
instrument
< 0.005 nm
Based on an
instrument
group
< 0.020 nm
Photometric noise
400–700 nm
< 50 µAU
700–2200 nm
N/A
700–2500 nm
< 20 µAU
850–1100 nm
N/A
1100–1600 nm
N/A
7.5
Reference conditions
Ambient tempera-
ture
+25 °C (±3 °C)
Relative humidity
≤ 60%
Instrument status
> 30 minutes in operation
Validity of the
data
After adjustment
7.6
Dimensions
Monochromator
Measurements
Width
380 mm
Height
348 mm
Depth
335 mm
Weight
21.0 kg
NIRS XDS RapidContent Analyzer module
Measurements
Width
380 mm
Height
351 mm
Depth
275 mm