5.2 Maintenance
■■■■■■■■■■■■■■■■■■■■■■
26
■■■■■■■■
NOTICE
Stabilizing time
The instrument features a monitoring function for stabilizing which
can be deactivated in the control software. If the monitoring is
deactivated, the instrument takes about 20 to 30 minutes for cor-
rect stabilizing.
11 Calibrate the instrument
The instrument has to be recalibrated every time a lamp is replaced.
■
Calibrate the instrument according to the specific procedure (see
tutorial for the control software).
The instrument is ready for operation again once the calibration and
the according tests have been carried out successfully.
5.2.4
Replacing the fuse
Replacing the fuse
1 Disconnecting the instrument from the power supply
■
Turn the on/off switch to the position
O
.
■
Pull out the power supply cable.
■
Remove the network cable (optional).
2 Removing the fuse holder
■
Pull out the fuse holder with a screwdriver.