3.1 Preparing the dialysis cell
■■■■■■■■■■■■■■■■■■■■■■
10
■■■■■■■■
IC equipment Dialysis 6.5330.100
■
Thoroughly rinse off the sealing ring
, the donor chamber
ultrapure water.
■
Dry all parts with a lint-free cloth.
5 Wetting the dialysis membrane
NOTICE
In the package containing the dialysis membranes, you will find
sheets of different thicknesses and colors:
■
The firm white cardboard is a cover protecting the filtration
membranes. Do not place it in the dialysis cell.
■
The thin light-blue sheets are separation sheets placed between
two filtration membranes. Do not place them in the dialysis cell.
■
The thin white sheets are the dialysis membranes. Use only
these for dialysis.
■
Using the tweezers, take a new dialysis membrane out of the
package.
■
Place the dialysis membrane in a petri dish filled with ultrapure
water and allow to hydrate for approx. two minutes.
6 Inserting the dialysis membrane
NOTICE
Make sure that the water-soaked dialysis membrane does not dry
out before it is inserted, as it can otherwise no longer be used.