3 Installation
IC equipment Dialysis 6.5330.100
■■■■■■■■
9
2 Removing the screws
■
Loosen the screws with the hex key.
■
Remove the screws with the washers and put them aside.
3 Disassembling the dialysis cell
■
Remove the upper part of the dialysis cell.
■
Remove the sealing ring.
4 Cleaning the dialysis cell
CAUTION
Damage to the dialysis cell
Organic solvents (e.g. acetone) corrode and damage the dialysis
cell material (PMMA).
■
Use only ultrapure water or a water-ethanol mixture (70:30) for
cleaning the dialysis cell.
■
For samples that contain organic components (e.g. solvents),
use the PEEK dialysis cell (6.2729.120). This cell has an excellent
chemical resistance to organic chemicals.