5.5 Connecting USB devices

■■■■■■■■■■■■■■■■■■■■■■

20

■■■■■■■■

915 KF Ti-Touch

5.5.6 

Connecting a barcode reader

The barcode reader is used as an aid for text and numerical input. You can
connect a barcode reader with USB interface.

Connect the barcode reader as follows:

 

1

Connect the USB plug of the barcode reader with the USB connector
of the Ti-Touch (Type A).

 

2

Enter and configure the barcode reader in the device manager.

Settings on the barcode reader:

Program the barcode reader as follows (also see manual for the barcode
reader):

1

Switch the barcode reader to programming mode.

 

2

Specify the desired layout for the keyboard (USA, Germany, France,
Spain, German-speaking Switzerland).

This setting must match the setting in the device manager.

 

3

Make sure that the barcode reader is set in such a way that Ctrl char-
acters (ASCII 00 to 31) can be sent.

 

4

Program the barcode reader in such a way that the ASCII character
02 (STX or Ctrl B) is sent as the first character. This first character is
normally referred to as the "Preamble" or "Prefix Code".

 

5

Program the barcode reader in such a way that the ASCII character
04 (EOT or Ctrl D) is sent as the last character. This last character is
normally referred to as the "Postamble", "Record Suffix" or "Postfix
Code".

 

6

Exit the programming mode.

5.5.7 

Connecting a Sample Processor

If you wish to integrate your Ti-Touch in an automation system, then you
can connect the following Sample Processors to the USB connector:

814 USB Sample Processor

815 Robotic USB Sample Processor XL

810 Sample Processor (from firmware version 5.915.0040)

Содержание 915 KF Ti-Touch

Страница 1: ...915 KF Ti Touch Manual Short Instructions 8 915 8004EN...

Страница 2: ......

Страница 3: ...Metrohm AG CH 9100 Herisau Switzerland Phone 41 71 353 85 85 Fax 41 71 353 89 01 info metrohm com www metrohm com 915 KF Ti Touch Manual Short Instructions 8 915 8004EN 04 2017...

Страница 4: ...tation has been prepared with great care However errors can never be entirely ruled out Please send comments regarding possible errors to the address above Documentation in additional languages can be...

Страница 5: ...disposal 8 4 Overview of the instrument 9 4 1 Front of the instrument 9 4 2 Rear of the instrument 10 5 Installation 11 5 1 Setting up the instrument 11 5 1 1 Packaging 11 5 1 2 Checks 11 5 1 3 Locati...

Страница 6: ...ecting a polarizable electrode 21 5 7 3 Connecting the temperature sensor 22 5 8 Connecting the Ti Touch to a network 23 6 Operation 25 6 1 Switching the instrument on and off 25 6 2 Fundamentals of o...

Страница 7: ...d manual In addition to an introduction safety instructions and an overview of the instruments you will also find information about the installation and operation of the 915 KF Ti Touch in addition to...

Страница 8: ...r You manage titrants sensors methods etc conveniently in the internal memory of the 915 KF Ti Touch You can also save your files externally through the USB connector e g on a USB flash drive On this...

Страница 9: ...ive or a USB hub Sensor connectors One connector each for Polarizable electrodes Temperature sensor Pt1000 or NTC Ethernet connector For connecting the Ti Touch to a network Two connection nipples For...

Страница 10: ...refers to the figure number the sec ond to the instrument part in the figure Instruction step Carry out these steps in the sequence shown Method Dialog text parameter in the software File New Menu or...

Страница 11: ...2 Introduction 915 KF Ti Touch 5 NOTE This symbol highlights additional information and tips...

Страница 12: ...e electrical safety when working with the instrument is ensured as part of the international standard IEC 61010 WARNING Only personnel qualified by Metrohm are authorized to carry out service work on...

Страница 13: ...ell by hand Avoid applying excessive force to tubing connections Damaged tubing ends lead to leakage Appropriate tools can be used to loosen connections Check the connections regularly for leakage If...

Страница 14: ...U WEEE Waste from Electrical and Electronic Equipment The correct disposal of your old instrument will help to prevent negative effects on the environment and public health More details about the disp...

Страница 15: ...help for the dialog dis played 5 Fixed key Print Opens the print dialog 6 Key Pressing the key pumps air into the solvent bottle The overpressure in the solvent bottle pushes solvent into the KF titr...

Страница 16: ...RJ 45 For connecting to a network 3 Temperature sensor connector Temp For connecting temperature sensors Pt1000 or NTC Two B sockets 2 mm 4 Electrode connector Pol For connecting polarizable electrode...

Страница 17: ...d may not be used in explosive environments Place the instrument in a location of the laboratory which is suitable for operation free of vibrations protected from corrosive atmosphere and contaminatio...

Страница 18: ...tected against accidental disconnection of the cable by means of a pull out protection fea ture If you wish to pull out the plug you will first need to pull back the outer plug sleeve marked with arro...

Страница 19: ...6 1415 250 KF titration cell with a 6 1903 020 or 6 1903 030 stirring bar inside on the vessel lid Fold back the holding bracket upwards while doing so The markings on the vessel lid and on the plasti...

Страница 20: ...8 pin Mini DIN socket and be control led simultaneously by the Ti Touch In addition to the connection cable stirrers and the Remote Box are each equipped with their own MSB socket for this purpose NO...

Страница 21: ...rking on the MSB connector see Figure 5 page 15 Figure 5 MSB connector CAUTION Switch off the Ti Touch before you plug in MSB devices When it is switched on the Ti Touch automatically recognizes which...

Страница 22: ...e controlled via remote lines and or that send control signals via remote lines can be connected via the 6 2148 010 Remote Box In addition to Metrohm other instrument manufacturers also use similar co...

Страница 23: ...the USB you can use a commercially avail able USB hub NOTE We recommend that the Ti Touch be switched off while you set up or disconnect connections between the devices 5 5 2 Connecting a USB hub Use...

Страница 24: ...interface Balance Cables AND ER FR FX with RS 232 inter face OP 03 6 2125 020 6 2125 010 Mettler AB AG PR LC RS9 In the scope of delivery for the balance Mettler AM PM PE with interface option 016 or...

Страница 25: ...for cable 3 Switch on the balance 4 If necessary switch on the RS 232 interface of the balance 5 Configure the RS 232 interface of the USB RS 232 adapter in the device manager of the Ti Touch 6 Enter...

Страница 26: ...t match the setting in the device manager 3 Make sure that the barcode reader is set in such a way that Ctrl char acters ASCII 00 to 31 can be sent 4 Program the barcode reader in such a way that the...

Страница 27: ...If you wish to pull out the plug you first need to pull back the outer plug sleeve marked with arrows 3 Configure the Sample Processor in the device manager 5 6 Setting up the titration vessel The tu...

Страница 28: ...le by means of a pull out protection If you wish to pull out the plug again you first need to pull back the outer plug sleeve 5 7 3 Connecting the temperature sensor A temperature sensor of the Pt1000...

Страница 29: ...cting the Ti Touch to a network The 915 KF Ti Touch has a network connection Ethernet This can be used to integrate your Ti Touch in your network You can for example store data on a PC within the netw...

Страница 30: ...5 8 Connecting the Ti Touch to a network 24 915 KF Ti Touch Figure 8 Connecting the Ti Touch to a network...

Страница 31: ...TE English is set as the default dialog language when the instrument is switched on for the first time Information about changing the dialog language can be found in the chapter Selecting the dialog l...

Страница 32: ...nnected then a request appears to carry out the Prepare function All tubings and the cylinder are rinsed with the Prepare function The preparing of the buret unit is described in the chapter Manual co...

Страница 33: ...2 Fundamentals of operation 6 2 1 Touch sensitive screen The entire 915 KF Ti Touch user interface is touch sensitive Simply touch a few of the buttons on the interface to learn how a touch sensitive...

Страница 34: ...dialog displayed Print opens the printing dialog Manual opens the manual control Stop cancels the running determination Start starts a determination The file name of the loaded method the time and the...

Страница 35: ...A check box can also be activated or deactiva ted by tapping on it 6 2 3 Status display The current status of the system is displayed in the upper right hand cor ner of the title bar Table 3 Status di...

Страница 36: ...ng the modification Delete entry The content of the input field is deleted com pletely The character in front of the cursor is deleted The cursor within the input field is shifted to the left by one c...

Страница 37: ...can be entered then the corresponding but tons will be shown to the right of the numerical keypad R1 For many parameters a result previously defined in the method can also be entered in place of a num...

Страница 38: ...tive regarding the precise terms and conditions involved in concluding a corresponding maintenance agreement For detailed information on this topic please visit www metrohm com Cleaning Cleaning the s...

Страница 39: ...ixed key 9 I Installation Set up instrument 11 Instrument Switch off 27 Switch on 25 K Keyboard Connect 19 KFT 2 M Magnetic stirrer Connect 16 Metrohm Serial Bus MSB see also MSB 14 MSB Connect device...

Отзывы: