7 Technical specifications
867 pH Module
■■■■■■■■
37
7.1.2
Temperature
A measuring input (
Temp.
) for temperature sensors of the Pt1000 or NTC
type with automatic temperature compensation.
R (25 °C) and B value can be configured for NTC sensors.
Measuring range
Pt1000
–150 - +250 °C
NTC
–5 - +250 °C
(R (25 °C) = 30000
Ω
and B (25/50) = 4100 K)
Resolution
Pt1000
0.1 °C
NTC
0.1 °C
Measuring accu-
racy
Pt1000
±0.2 °C
(Applies for measuring range –20 - +150 °C; ±1 digit; without sensor
error, under reference conditions)
NTC
±0.6 °C
(Applies for measuring range +10 - +40 °C; ±1 digit; without sensor
error, under reference conditions)
7.1.3
Polarizer
One measuring input (
Pol.
) for polarizable electrodes.
Measuring mode
Ipol
Determination with adjustable polarization current
Polarization cur-
rent
–122.5 - +122.5 µA (increment: 0.5 µA)
–125.0 - +125.0 µA: non-guaranteed values, dependent on reference
v2.5 mV
Measuring
range
–1200 - +1200 mV
Resolution
0.1 mV
Measuring
accuracy
±0.2 µA
(±1 digit; without sensor error, under reference conditions)
Measuring mode
Upol
Determination with adjustable polarization voltage
Polarization
voltage
–1,225 - +1,225 mV (increment: 25 mV)
–1,250 - +1,250 mV: non-guaranteed values, dependent on reference
v2.5 mV
Measuring
range
–120 - +120 µA