1. With the POWER switch OFF, plug the supply power cord into an appropriate,
grounded, receptacle. (Refer to the cabinet data plate for voltage and
ampere rating).
2. Ensure that the cabinet’s floor duct and water pan are installed. (refer to
the cabinet operating instructions). This helps to distribute the heat evenly
throughout the cabinet.
3. If extra humidity is desired, fill the water pan in the floor duct of the cabinet
with clean warm water. The capacity of the pan is about one quart (one liter).
4. Set the POWER switch to the ON position. The red power light next to the
switch will glow when the power is switched ON and will continue to glow until
switched OFF.
5. Set the TEMPERATURE control to 6 or 7. This should provide a temperature
of 150° to 170°F (66° to 77°C) after allowing the cabinet to preheat for 45 to
60 minutes.
6. The red indicator light between the thermostat and thermometer will turn on
and off as the thermostat and heat elements cycle.
7. To change temperature set point, turn the TEMPERATURE control knob.
About 30 minutes after a knob is turned, the thermometer should accurately
display the change in cabinet temperature.
8. To clean cabinet, turn power off, unplug from wall outlet, allow the cabinet to
cool, remove the floor duct and pan, wipe the cabinet with a damp cloth and
dry with a towel.
9. Consult the instructions for use for additional operation and maintenance
information.