Technical characteristics of the MTX 3290
42
6,000- and 60,000-count digital multimeters
Technical characteristics of the MTX 3290 (continued)
Currents
Three possible modes: DC, AC, AC+DC
In DC mode, you can measure a direct current or the DC component of an
alternating current.
In the AC and AC+DC modes, you can measure the true RMS (TRMS) value of
an alternating current with/without its direct component (no capacitive coupling in
"DC" mode).
DC current
Particular reference conditions:
6mA range: Measuring a strong current for a long time can cause a temperature
rise in some components. In this case, it is necessary to wait some time for the
metrological characteristics specified in 6mA to be restored.
Range
Operating
range
Specified
measurement
range
Resolution
Uncertainty
(
±
)
Voltage
drop
Protection
6mA
0 to 6.000mA
0.002 to
6.000mA
1µA
1.2% L ± 5D
25mV/mA
Fuse
10A/600V
>50kA
60mA
0 to 60.00mA 0.02 to 60.00mA
0.01mA
1.2% L ± 2D
3mV/mA
600mA
0 to 600.0mA 0.2 to 600.0mA
0.1mA
1.2% L ± 2D 0.58mV/mA
6A
0 to 6.000A 0.200 to 6.000A
0.001A
1.2% L ± 3D
0.05V/A
10 A/20A
(∗)
0 to 20.00A
0.20 to 20.00A
0.01A
1.2% L ± 2D
0.05V/A
The display indicates "OL" above 19.99A. The
symbol blinks and a beep
sounds above 10A.
(∗)
Acceptable overload: 10A to 15A for 30s max. with a pause of 5min between
2 measurements. Ambient temp. 35°C max.
Secondary measurements and displays: MAX, MIN, AVG
A
AC RMS current
Range
Operating
range
Specified
measurement
range
Resolution
Uncertainty (
±
)
40Hz to 20kHz
(
**
)
Peak factor
Voltage
drop
Protection
6mA
0 to
6.000mA
0.600 to
6.000mA
1µA
1.7% L ± 5D
2.6 to
5mA
25mV/mA
Fuse
10A/600V
>50kA
60mA
0 to
60.00mA
6.00 to
60.00mA
0.01mA
1.5% L ± 3D
2.6 to
50mA
3mV/mA
600mA
0 to
600.0mA
60.0 to
600.0mA
0.1mA
2.6 to
500mA
0.58mV/mA
6A
0 to
6.000A
0.600 to
6.000A
0.001A
1.7% L ± 5D
2.8 to 5A
0.05V/mA
10A/20A
(∗)
0 to
20.00A
1.00 to
10.00A
0.01A
1.5% L ± 3D
3.7 to 8A
0.05V/mA
The display indicates "OL" above 19.99A. The
symbol blinks and a beep
sounds above 10A.
Secondary measurements and displays: FREQ (AC coupling) MAX, MIN, AVG,
PEAK
(∗)
Acceptable overload: 10A to 15A for 30s max. with a pause of 5min between 2
measurements. Ambient temp. 35°C max.
(∗∗)
Additional uncertainty with the 300Hz filter.