MI 2892 Power Master
Operating the instrument
66
Table 3.55: Keys in General recorder setup screen
START
STOP
Starts the recorder.
Stops the recorder.
HELP
Show help screen where it’s explained which measurements will be
recoded with Limited and Standard profile.
See section 4.4 for details.
CONFIG
Shortcut to Connection setup. See 4.2 for details.
CHECK C.
Check connection settings. See 3.21.1 for details.
Enters recorder starting date/time setup.
Keys in Set start time window:
Selects parameter to be changed.
Modifies parameter.
Confirms selected option.
Exits Set start time window without modifications.
Selects parameter to be changed.
Modifies parameter.
Returns to the “RECORDERS” submenu.
3.15
Waveform/inrush recorder
Waveform recording is a powerful tool for troubleshooting and capturing current and voltage
waveforms and inrushes. Waveform recorder saves a defined number of periods of voltage and current
on a trigger occurrence. Each recording consists of pre-trigger interval (before trigger) and post-trigger
interval (after trigger).
F1
F2
F3
F4
ENTER
ENTER