MI 2892 Power Master
Theory and internal operation
178
voltage U
Rms(1/2)
values (one cycle RMS voltage refreshed each half cycle) is measured for each interval.
In following figures, Level triggering is shown.
Inrush, fluctuation or other event
Measured signal
Inrush record
t
U or I
t
I
R
ms(1/2)
or
U
Rms(1/2)
Trigger Level
Slope: Fall
U or I
Trigger point
Figure 5.25: Level triggering
Triggering slope
S
lo
p
e
:
r
is
e
t
t
S
lo
p
e
:
f
a
ll
Figure 5.26: Triggering slope
5.1.20
Transient recorder
Transient recorder is similar to waveform recorder. It stores a selectable set of pre- and post-trigger
samples on trigger activation, but with 10 times higher sampling rate.
Recorder can be triggered on envelope or level.
Envelope trigger is activated if difference between same samples on two consecutive periods of
triggering signal, is greater than given limit.