MD 9235 Clamp Meter Series
Specifications
17
5 Specifications
Display :
Voltage functions:
6000 counts LCD display
Power, Ohm & Hz functions:
9999 counts LCD display
ACA clamp-on function:
4000 counts LCD display
Update Rate :
Power function:
2 per second nominal
Voltage, ACA clamp-on & Ohm functions: 2 per second nominal
Hz function:
1 per second nominal
Polarity :
Automatic
Low Battery :
Below approx. 2.4V
Operating Temperature :
0
o
C to 40
o
C
Relative Humidity :
Maximum relative humidity 80% for emperature up
to 31
o
C decreasing linearly to 50% relative
humidity at 40
o
C
Altitude :
Operating below 2000m
Storage Temperature :
-20
o
C to 60
o
C, < 80% R.H. (with battery removed)
Temperature Coefficient :
nominal 0.15 x (specified accuracy)/
o
C @(0
o
C -
18
o
C or 28
o
C -40
o
C), or otherwise specified
Sensing :
True RMS sensing
Safety :
Meets IEC/UL/EN61010-1 Ed. 3.0, IEC/EN61010-
2-032 Ed. 3.0, IEC/EN61010-2-033 Ed. 1.0,
IEC/UL/EN61010-031 Ed. 1.1 & CAN/CSA-C22.2
No. 61010-1-12 Ed. 3.0
Measurement Category:
III 600 Volts ac & dc
Transient protection:
6.5kV (1.2/50
s surge)
Pollution degree:
2
E.M.C.:
Meets EN61326-1:2006 (EN55022, EN61000-3-2,
EN61000-3-3,
EN61000-4-2,
EN61000-4-3,
EN61000-4-4,
EN61000-4-5,
EN61000-4-6,
EN61000-4-8, EN61000-4-11)
In an RF field of 3V/m:
Total Accuracy = Specified Ac 50 digits
Performance above 3V/m is not specified
Overload Protections :
ACA Clamp-on jaws :
AC 600A rms continuous
+ & COM terminals (all functions) : 600VDC/VAC rms
Power Supply:
standard 1.5V AAA Size (NEDA 24A or IEC LR03)
battery X 2
Power Consumption :
Voltage, ACA, Hz & Power functions: 11mA typical
Ohm function:
5.5mA typical
APO Timing :
Idle for 30 minutes
APO Consumption :
4
A typical
Dimension :
L189 X W78 X H40 mm
Weight :
192 gm approx
Jaw opening & Conductor diameter :
26mm max