MD 9235 Clamp Meter Series
Specifications
19
ACA AutoVA
TM
Threshold: 1A AC (40Hz ~ 500Hz only) nominal
Crest Factor:
< 3 : 1 at full scale & < 6 : 1 at half scale for 40.00A, 400.0A & 600A ranges
1)
Induced error from adjacent current-carrying conductor: < 0.06A/A
2)
Specified accuracy is from 1% to 100% of range and for measurements made at the
jaw center. When the conductor is not positioned at the jaw center, position errors
introduced are:
Add 1% to specified accuracy for measurements made WITHIN jaw marking lines (away
from jaw opening)
Add 4% to specified accuracy for measurements made BEYOND jaw marking
lines (toward jaws opening)
PEAK-rms HOLD (ACA & ACV only)
Response: 65ms to >90%
Frequency
RANGE
Accuracy
5Hz ~ 500Hz
0.5%+4d
Sensitivity (Sine RMS)
40A range: > 4A
400A range: > 40A
600A range: > 400A
600V range: > 30V
Ohms
RANGE
Accuracy
999.9
1.0% + 6d
Open Circuit Voltage : 0.4VDC typical
Audible Continuity Tester
Audible threshold: between 10
and 300
.
Response time: 250
s
Single-Phase & 3-Phase Balanced-Load Power
RANGE
5)
0 ~ 600.0kVA
Accuracy
1) 2) 3)
F ~ 10th
11th ~ 45th
46th ~ 51st
@ PF = 0.99 ~ 0.1
2.0%+6d
3.5%+6d
5.5%+6d
RANGE
5)
0 ~ 600.0kW / kVAR
Accuracy
1) 2) 4)
F ~ 10th
11th ~ 25th
26th ~ 45th 46th
~
51st
@ PF = 0.98 ~ 0.70
2.0%+6d
3.5%+6d
4.5%+6d
10%+6d
@ PF = 0.70 ~ 0.50
3.0%+6d
@ PF = 0.50 ~ 0.30
4.5%+6d
@ PF = 0.30 ~ 0.20
10%+6d
15%+6d
1)
Specified accuracy is for ACA clamp measurement at the center of jaws. When the
conductor is not positioned at the jaw center, position errors introduced are:
Add 1% to specified accuracy for ACA measurements made WITHIN jaw marking lines
(away from jaw opening)