MD 9010 General Purpose Auto – Check Digitalmultimeter
Specification
10
6 Specification
GENERAL SPECIFICATIONS
Display & Update Rate:
3-5/6 digits 6000 counts; Updates 5 per second nominal
Operating Temperature:
0
o
C ~ 40
o
C
Relative Humidity:
Maximum relative humidity 80% for temperature up to 31
o
C decreasing linearly to
50% relative humidity at 40
o
C
Altitude:
Operating below 2000m
Storage Temperature:
-20
o
C ~ 60
o
C, < 80% R.H. (with battery removed)
Temperature Coefficient:
Nominal 0.15 x (specified accuracy)/
o
C @ (0
o
C ~ 18
o
C or 28
o
C ~ 40
o
C), or
otherwise specified
Sensing:
Average sensing
Pollution Degree:
2
Safety:
Meets IEC61010-1, UL61010B-1, CAN/CSA-C22.2 No. 1010.1-92
Transient Protection:
4kV lightning surge (1.2/50
s)
Measurement Category:
CAT II 600V and CAT III 300V
E.M.C.:
Meets EN61326 (1997, 1998/A1), EN61000- 4-2 (1995), & EN61000-4-3 (1996)
In an RF Field of 3V/m:
Capacitance function is not specified
Other function ranges:
Total accuracy = Specified ac 45d
Performance above 3V/m is not specified
Overload Protection:
600VDC & VAC rms
Low Battery:
Below approx. 2.4V
Power Supply:
3V standard button battery x 1 (IEC-CR2032; ANSI-NEDA-5004LC)
Power Consumption (typical):
6mA for Voltage functions on Auto-V
position, and 2mA for other functions
APO Consumption (typical):
2.2
A
APO Timing:
Idle for 3 minutes
Dimension / Weight:
L113mm x W53mm x H10.2mm / Approx. 78 gm
Special Features:
AutoCheck
TM
(Automatic V &
Selection), and EF-Detection
Accessories:
Battery installed, and User’s manual
Electrical Specification
Accuracy is given as +/- (% of reading number of digits) or otherwise specified @ 23
o
C +/- 5
o
C
and less than 75% R.H.