MI 2087 AL2 EasiTEST
Measurement instructions
17
Fig. 10:
Connection of optional Tip Commander (Order No.
A 1002
) and optional
Probe Test Lead (Order No.
A 1012
)
Step 5
q
Press the
START
key and release it. Measurement will be carried out and
result displayed afterwards. Each measurement is accomplished in two steps
(polarity is reversed between the two steps automatically). Higher partial result
is displayed.
q
Check the lower partial result by pressing the
DISPLAY
key.
Notes:
q
In case of present external voltage higher than 9V (AC or DC) between test
terminals, the continuity measurement will not be carried out after pressing
START
key instead the voltage will be displayed, equipped with
“
!
” mark!
Warning sound is also emitted.
q
If resistance value higher than 5
Ω
(measured with not-compensated
instrument) is displayed, compensation will not be carried out after pressing
CAL
key, but already effected compensation will be annulled (“
C
ü
” mark will
disappear)!
q
If test result is out of measurement range (open test leads) “
>1999
Ω
” message
will be displayed!
q
“
bat
” message, displayed during or after finishing the measurement means
batteries are too weak to guarantee correct result. Replace the batteries.