MI 2087 AL2 EasiTEST                                                       

Measurement instructions 

 

16 

 

Step 2 

 

Compensate test leads 

(if they have not been compensated yet or, if already 

compensated test leads have been exchanged) as follows:

 

 

Short test leads; see the figure below. 

 
 

 

Fig. 8:

 Shorted test leads

 

 
 

Press the 

START

 key and release it in order to carry out regular measurement, 

result (close to 0 

) will be displayed. 

Press the 

CAL

 key and release it, displayed result will alter to 

0,00

 

and 

C

ü

 

mark will appear, indicating the compensation was successfully accomplished. 
Test instrument is thus ready to be used. 

In order to remove potential compensation, follow the procedure described in 

this step with open test leads. “

C

ü

 

mark will disappear after finishing the 

procedure, indicating the compensation has been removed. 

 
 
 

Step 3 

 

Connect test leads to the tested object according to the figures below. 

 
 

 

Fig. 9:

 Connection of Universal Test Cable and optional Probe Test Lead  

(Order No. 

A 1012

)

 

N/L2

L/L1

L/L1

PE/L3

N/L2

Opt

ion

 A 1

002

PE/L3

PCC1

Option A 1012

PCC2

PCC3

N/L2

PE/L3

L/L1

MPEC

MPEC..........Main Potential 
                     Equilizing Collector
PCC.............Protection Conductor
                     Collector 

Содержание EasiTEST MI 2087 AL2

Страница 1: ...EasiTEST MI 2087 AL2 Instruction Manual Version 1 Code No 20 750 810...

Страница 2: ...el metrel si Mark on your equipment certifies that this equipment meets the requirements of the EU European Union concerning safety and interference causing equipment regulations 2003 METREL No part o...

Страница 3: ...3 Measurement instructions 13 3 1 Insulation resistance 13 3 2 Continuity of protective conductors 15 3 3 RCD Contact voltage and fault loop resistance 18 3 4 RCD Trip out time 20 3 5 RCD Trip out cu...

Страница 4: ...well as wide range of subresults parameters and messages Operation is simple and clear operator does not need any special training except to read this instruction manual to operate the instrument In o...

Страница 5: ...SC 0 264 V Line resistance RL N LOOP L N PSC Max test current 2 5 A Line prospective short circuit current Ipsc LOOP L N PSC Calculation Ipsc UN 1 06 RL N Voltage UL N LOOP L N PSC 0 264 V RCD Contact...

Страница 6: ...rds applied Instrument operation IEC EN 61557 1 IEC EN 61557 2 IEC EN 61557 3 IEC EN 61557 4 IEC EN 61557 6 DIN VDE 100 BS 7671 16th edition Electromagnetic compatibility EMC EN 61326 Safety IEC EN 61...

Страница 7: ...MI 2087 AL2 EasiTEST Instrument description 7 2 Instrument description 2 1 Front panel 1 2 11 10 9 8 7 6 5 4 3 Fig 1 Front panel...

Страница 8: ...est result is displayed q Check Fault Loop Resistance value and set Limit Contact Voltage TRIP LOCK position when result Uc is displayed q Check Contact Voltage and set Limit Contact voltage RCD MANUA...

Страница 9: ...description 9 2 2 Connector panel 1 Use original test cables only Max allowed voltage between test terminals and ground is 300V Max allowed voltage between test terminals is 500V Fig 2 Connector pane...

Страница 10: ...trument around operator body 3 Plastic cover it fixes nylon strip to the instrument There is a screw under the cover which has to be unscrewed when opening the instrument for service or calibration pu...

Страница 11: ...tes of carring the instrument are available Operator can choose appropriate one for the intended operation see the following examples The instruments is hung around operator s neck only quick placing...

Страница 12: ...et configuration or contact your dealer for further information FUNCTION REQUIRED ACCESSORIES Insulation resistance Universal Test Cable or Tip Commander Option Order No A 1002 Continuity of protectiv...

Страница 13: ...ed before starting the measurement q When measuring insulation resistance between conductors all loads must be disconnected and all switches closed q Do not touch tested object while testing it risk o...

Страница 14: ...ter setting nominal test voltage to return to Insulation resistance initial menu Step 3 q Connect test cable to the tested object according to the figure below Fig 6 Connection of Universal Test Cable...

Страница 15: ...er finishing the measurement means batteries are too weak to guarantee correct result Replace the batteries 3 2 Continuity of protective conductors Continuity of protective conductors has to be measur...

Страница 16: ...appear indicating the compensation was successfully accomplished Test instrument is thus ready to be used q In order to remove potential compensation follow the procedure described in this step with...

Страница 17: ...than 9V AC or DC between test terminals the continuity measurement will not be carried out after pressing START key instead the voltage will be displayed equipped with mark Warning sound is also emit...

Страница 18: ...pears Fig 11 Contact voltage initial menu left and Fault loop resistance initial menu Step 2 q Select required function i e Contact voltage or Fault loop resistance by pressing the SELECT key first Pr...

Страница 19: ...sing the and keys 50 V or 25 V can be selected q Press the SELECT key again after selecting the limit contact voltage to return to Contact voltage or Fault loop resistance initial menu Step 6 q Connec...

Страница 20: ...PE terminals will influence the test result Disconnect such appliances before the measurement is started q Specified accuracy of Uc RL measurement is valid only if earthing system connected to PE term...

Страница 21: ...inal differential current as follows q Press the SELECT key after setting nominal differential current last set multiplier starts to blink q Select appropriate value by using the and keys The followin...

Страница 22: ...measurement is carried out otherwise the test is not relevant That is why waiting time of 30s is involved before Trip out Time measurement The waiting time is presented on display as countdown from 3...

Страница 23: ...ential current starts to blink q Select appropriate value using the and keys The following values are available to be selected 10mA 30mA 100mA 300mA 500mA or 1000mA Step 3 q Connect test cable to test...

Страница 24: ...ied accuracy of tested parameters is valid only if earthing system connected to PE terminal is free of interfering voltages q bat message displayed during or after finishing the measurement means batt...

Страница 25: ...ype S is displayed can be selected Step 4 q Connect test cable to tested object mains outlet or other test terminals according to the figure 12 Universal test cable Step 5 q Press the START key and re...

Страница 26: ...will appear for a while Successfully ended third test RCD trips within allowed time period will be followed by rcd message informing the operator to reswitch the RCD see the figure below right side F...

Страница 27: ...wed by rcd message informing the operator to reswitch the RCD see the figure below right side Fig 21 Fifth test is running left figure rcd message displayed after tripping out RCD right figure After r...

Страница 28: ...re Trip out Time measurement is carried out otherwise the test is not relevant That is why 30 s of waiting time is involved in some stepsm namely Step 3 Step 4 Step 5 and Step 6 The waiting time is pr...

Страница 29: ...e Universal test cable to EasiTest q Set function switch to LOOP L E PSC position L PE voltage measurement will start to run see the figure below left side Fig 23 L PE voltage and frequency measuremen...

Страница 30: ...nals automatically if N L2 and L L1 test leads Universal test cable are connected reversed way or terminals on tested wall plug are reversed q If test result is out of display range 1999 message will...

Страница 31: ...to the figure 24 phase to neutral measurement or according to the figure below phase to phase measurement Fig 26 Phase to phase measurement Step 3 q Press the START key and release it Measurement wil...

Страница 32: ...played after pressing START key Warning sound is also emitted q If test result is out of display range 1999 message will be displayed q Specified accuracy of tested parameters is valid only if mains v...

Страница 33: ...rument q Switch OFF the instrument q Press the CAL key and keep it pressed while switching ON the instrument rES message will be displayed for a while meaning RESET function has been completed Paramet...

Страница 34: ...hey are empty in static mode no measurement is running or when the battery indicator shows an empty frame just temporarily while a measurement is running Nominal power supply voltage is 6 VDC Use four...

Страница 35: ...e fuses F2 and F3 if one of the following malfunctions is noticed q rcd message is blinking after pressing the START key in all RCD functions as well as in RLINE Ipsc and RLOOP Ipsc function q 1999 me...

Страница 36: ...calibration is carried out every sixth months 5 5 Service Repairs under or out of warranty time Please contact your distributor for further information Distributor s address Producer s address METREL...

Страница 37: ...at 0 or 180 RCD type standard or selective Contact voltage Uc Measuring range 10 100 V Display range V Resolution V Accuracy 0 00 9 99 0 01 0 10 of r 0 2 V 10 0 100 0 0 1 0 10 of r The accuracy is val...

Страница 38: ...1999 Display range Resolution Accuracy 0 00 19 99 0 01 20 0 199 9 0 1 200 1999 1 5 of r 5 D Prospective short circuit current Ipsc Display range A Resolution A Accuracy 0 06 19 99 0 01 20 0 199 9 0 1...

Отзывы: